Plug-in control
53
GA 05.141/6.02 - 07/2003
Start
Stop
Esc
Mon
Prog
Enter
+
_
Acceleration
14.5A 254Hz HOK
Key
Function
Esc
• Returns to the operating display from the storage procedure
without storage.
• Returns to the operating display from any point of the basic
menu.
Mon
• No function
Prog
• Selects the programming menu from the operating display.
• Confirms to store changed parameters to the EEPROM
Enter
• Switches forward to the next submenu
Start
• Starts the pump (only possible if there is no fault).
The start key is only active if the user is in the basic menu
or in the operating display.
Stop
• Stops the pump (only from the basic menu)
• Returns to the operating display from the programming menu.
• Acknowledges a failure after the cause of the failure has been
removed (only possible, if the user is in the operating display).
+
• Increases a parameter value or proceeds to the next option.
–
• Lowers a parameter value or returns to the previous option.
<
• Selects the programming menu from the operating display.
• Switches back to the last main menu.
>
• Selects the programming menu from the operating display.
• Switches forward to the next main menu.
Fig. 43 Functions of the front panel keys
Acceleration
14.5A 254Hz HOK
Operating display
Operating status
Motor current / Actual frequency / TMS status code*
If a critical operating status occurs this warning is
displayed alternating with the operating display.
Failure
Accel. Time
Failure message
Failure cause
TMS status code*
No.
Code Description
——————————————————————————
1
H
Heating pump
2
H O K
Temp. TMS ok, heater ON
3
O K
Temp. TMS ok
4
C O K
Temp. TMS ok, cooling ON
5
W
Temp. TMS > TMS
Set
+5K
O F F
TMS cancelled via control plug X14
6
H C O K
Temp. TMS ok, cooling ON, heater ON
7
N O K
Temp. TMS not ok, cooling OFF, heater OFF
8
H C
Temp. TMS not ok, cooling ON, heater ON
9
C
Temp. TMS not ok, cooling ON
see also Section 4.4 “TMS”
* only for pumps with TMS
Fig. 44 Display