background image

levr

 

LIM-05 user manual 

Dated 25/5/2017 

http://www.levr.pl/

 

Page 

4

 of 

20

   

 

1

1

.

.

 

 

P

P

r

r

e

e

i

i

n

n

s

s

u

u

l

l

a

a

t

t

e

e

d

d

 

 

d

d

i

i

s

s

t

t

r

r

i

i

c

c

t

t

 

 

h

h

e

e

a

a

t

t

i

i

n

n

g

g

 

 

p

p

i

i

p

p

e

e

l

l

i

i

n

n

e

e

 

 

a

a

s

s

 

 

a

a

 

 

t

t

e

e

s

s

t

t

e

e

d

d

 

 

o

o

b

b

j

j

e

e

c

c

t

t

 

 

Production of preinsulated pipes with sensor wires ensures good reproducibility of the 
product  both  in  terms  of  the  physical  properties  of  materials  and  reliable  geometrical 
dimensions.  This  ensures  that  certain  electrical  properties  are  reproduced  in  all 
products.  One  of  the  electrical  quantities  characteristic  to  district  heating  pipelines  is 
characteristic  impedance,  amounting  to  approximately  200 

.  The  impedance  is 

measured between the carrier pipe and the copper sensor wire. It should be emphasised 
that  impedance  does  not  depend  on  the  length  and  diameter  of  the  district  heating 
pipeline.  However,  it  may  change  due  to:  local  change  of  the  distance  between  the 
copper  wire  and  the  steel  pipe,  appearance  of  moisture  in  polyurethane  insulation, 
incorrect connection or a break in the detector line. These cases are regarded as faults 
that  should  be  located  and  eliminated.  Faults  are  located  using  a  reflectometer. 
Measurement with this instrument requires generation of a very short electric pulse. It 
is  transmitted  between  the  copper  wire  and  the  steel  pipe.  Since  the  characteristic 
impedance of the measured circuit is constant, the electric pulse moves freely along the 
tested pipeline with speed almost equal to the speed of light propagation. If one of the 
indicated  faults  occur  at  some  point  of  the  preinsulated  district  heating  pipeline, 
resulting in the abrupt change of characteristic impedance, the measurement pulse will 
be fully or partially reflected. The reflectometer measures the time from the sending of 
the  electrical  signal  until  its  reflection  returns.  Then,  using  the  entered  value  of 
measurement pulse speed and the measured time, it calculates the distance between the 
measurement station and the point of reflection, i.e. location of the fault. 
The method of fault location in preinsulated district heating pipelines with a detector 
loop described above is implemented using the LIM-05 instrument. 

2

2

.

.

 

 

G

G

e

e

n

n

e

e

r

r

a

a

l

l

 

 

i

i

n

n

f

f

o

o

r

r

m

m

a

a

t

t

i

i

o

o

n

n

 

 

a

a

b

b

o

o

u

u

t

t

 

 

t

t

h

h

e

e

 

 

L

L

I

I

M

M

-

-

0

0

5

5

 

 

i

i

n

n

s

s

t

t

r

r

u

u

m

m

e

e

n

n

t

t

 

 

The  LIM-05  fault  locator  is  a  stationary  instrument.  It  can  be  used  to  monitor  four 
sections of the district heating pipeline with a pulse alarm system at the same time. The 
monitored  section  of  the  sensor  wire  should  not  be  longer  than  2500 m.  The 
measurement  result  is  presented  on  a  graphical  display  in  the  form  of  an  annotated 
chart.  The  result  is  created  based  on  the  difference  between  the  previously  recorded 
reference run and the chart depicting the current condition. Until a fault is detected, the 
chart  is  a  straight  line  (axis x).  Axis x  is scaled  in  units  of  length  [m].  The  reference 
runs  for  each  monitored  section  of  the  network  are  recorded  automatically  in  the 
digital  memory  of  the  instrument  shortly  after  the  operator  initiates  the  measurement 
procedure. During initial calibration of the instrument, the length of axis x on the chart 
is adjusted to be a precisely proportionate representation of the length of the monitored 
section  of  the  district  heating  pipeline.  Information  about  the  fault,  in  turn,  has  the 
form  of  a  representation  of  the  reflected  measurement  pulse  depicting  its  actual 
electrical  shape.  The  location  (above  or  under  axis  x)  of  the  reflected  pulse  and  its 
shape indicate the type of the fault (e.g. leak, electrical short circuit of the sensor wire 
with  the  steel  pipe)  and  its  intensity.  The  intensity/degree/size  of  the  fault  is 
represented  by  the  amplitude  (height)  of  the  reflected  pulse.  The  amplitude  is 
measured in units of voltage [mV] and [V]. Axis y is scaled in these units by default. 
The distance from the beginning of axis x (x = 0) to the beginning (front) of the pulse 
is  proportionate  to  the  distance  between  the  beginning  of  the  district  heating  pipeline 

Содержание LIM-05

Страница 1: ...LIM 05 pulse fault locator user manual Software version 17 05 USER MANUAL EZUP LEVR s c 03 193 Warsaw ul Krzy wki 5 tel fax 48 22 639 85 66 Ver 17 12...

Страница 2: ...0 FN key 9 4 11 ENT key 10 4 12 ESC key 10 4 13 key 10 4 14 and keys 10 5 Main menu 10 5 1 Kana Channel option 10 5 2 Start Stop option 10 5 3 Kalibracja Calibration screen 10 5 4 Historia History opt...

Страница 3: ...16 Calibration procedure Kalibracja Calibration option D ugo p tli Loop length entry 16 Channel number entry 18 Initiate or stop automatic measurements 18 Creating and recording the reference run 18...

Страница 4: ...t calculates the distance between the measurement station and the point of reflection i e location of the fault The method of fault location in preinsulated district heating pipelines with a detector...

Страница 5: ...epicts the actual distance in m Figures 1 and 2 depict the method of presentation of measurement information Fig 1 Depiction of the reflected measurement pulse on the detector screen in case of a leak...

Страница 6: ...ength of the section of the alarm wire between these cursors x default axis x y default axis y Up Up voltage thresholds for initiation of automatic fault detection and indication the dead area is 2Up...

Страница 7: ...he operating conditions return to normal 3 1 Connection of the locator to the alarm system of the district heating pipeline The locator is connected to the carrier pipe and copper sensor wire with a c...

Страница 8: ...nment and the speed of light propagation in vacuum For district heating pipelines with a copper sensor wire this coefficient adopts values of the 0 91 0 93 range 4 2 2 zoomH key Press the 2 zoomH key...

Страница 9: ...of the screen after pressing the 6 K2 key 4 7 7 PING key Activating the PING function will force a measurement cycle in the selected measurement channel and display the measured run Depending on the...

Страница 10: ...ia Reference run 8 Info 5 1 Kana Channel option This option can be used to display measurement information about one of the monitored sections of the district heating pipeline To select the channel en...

Страница 11: ...iowy Input run option enables presentation of the collected digital data in the form of a chart These data refer to the last measurement When the input run is displayed the screen will display a descr...

Страница 12: ...ion method also ensures that the image corresponds to reality For instance if a fault occurs near the end of the network it will be represented near the end of the x axis and not for instance in the c...

Страница 13: ...ge value of measured results i which will be recorded as the reference run See section 0 6 5 Czas martwy Dead time entry The automatic measurement cycle of the fault locator consists of measurements c...

Страница 14: ...nitial stages of its development The shape of the reflected pulse will be represented within the range of voltage thresholds determining the measuring range Extending the voltage measuring range will...

Страница 15: ...ce of this message means that there is no communication between the computer and the locator In such a situation make sure that the devices are correctly connected and check the status of the selected...

Страница 16: ...ed only after the locator screen displays the run Calibration procedure Kalibracja Calibration option D ugo p tli Loop length entry Note Locator calibration should be carried out 30 minutes after the...

Страница 17: ...he section of the x axis visible on the chart does not reflect the actual length of the alarm wire In section 3 increased length of the line was specified precisely in order to ensure that the image o...

Страница 18: ...on of automatic measurements is indicated by the AUTO description It will appear in the bottom right corner of the screen Stopping automatic measurements will cause the description to disappear The se...

Страница 19: ...1 10 0 M Ma ai in nt te en na an nc ce e o of f t th he e L LI IM M 0 05 5 i in ns st tr ru um me en nt t A clean dry cloth should be used to remove dust from the housing of the instrument Other dirt...

Страница 20: ...range 2047 mV Maximum measurement sensitivity 1 mV pixel Sensitivity threshold voltage range 1 1024 mV Dead time range 1 255 min Methods of communication with external devices Alarm output volt free r...

Отзывы: