User Manual for BPS-i600 

www.levitronix.com 

 

PL-4089-00, Rev00, DCO# 21-271

 

 

First Release: 14-Dec-2019 

Last Update: 03-Dec-2021

 

 

 

 

BPS-i600 

3.1 bar              (45 psi) 
75 liters/min        (20 gallons/min) 

 
 

USER MANUAL

 

 
 

 

 
 
 
 

This manual contains information necessary for the safe and proper use of the 

BPS-i600

. Included are specifications 

for  the  standard  configurations  of  the  pump  system  and  instructions  regarding  its  use,  installation,  operation, 
adjustment,  inspection  and  maintenance.  For  special  configurations  of  the  pump  system  refer  to  accompanying 
information.  If  the  system  must  be  configured  for  other  parameter  settings,  then  the 

Levitronix

®

  Service  Software 

version

 V2.0.5.0 

or higher

 

(with according manual 

Levitronix

 Doc.# 

PL-4043-00

) is necessary. Familiarize yourself 

with the contents of the manual to ensure the safe and effective use of this product. After reading this manual, please 
store the manual where the personnel responsible for operating the pump system can readily refer to it at any time.

 

 

Содержание 100-91405

Страница 1: ...use installation operation adjustment inspection and maintenance For special configurations of the pump system refer to accompanying information If the system must be configured for other parameter settings then the Levitronix Service Software version V2 0 5 0 or higher with according manual Levitronix Doc PL 4043 00 is necessary Familiarize yourself with the contents of the manual to ensure the s...

Страница 2: ...NSTALLATION 16 4 1 Electrical Installation 16 4 1 1 Overview Wire Designation of Driver Cable 16 4 1 2 Overview Electrical Schematics of Driver Interface 17 4 1 3 Installation Instructions for Power Supply 18 4 1 4 Installation PLC Interface Signals 18 4 1 5 Installation of RS485 Interface 18 4 1 6 Installation of RS485 Fieldbus Multi Pump System Arrays 19 4 2 Mechanical Installation 19 4 2 1 Mech...

Страница 3: ... of sensitive electronic devices and measuring instruments in the immediate surroundings Keep at a safe distance from computers monitors and all magnetic data storage media e g disks credit cards audio and video tapes etc WARNING Hazardous voltage may be present In case of the usage of an inadequate AC DC power supply mains voltages may be present even if the system is designed for 48 VDC The use ...

Страница 4: ...al for BPS i600 www levitronix com PL 4089 00 Rev00 DCO 21 271 4 2 Specifications 2 1 Specification of Components Figure 1 Pump system with main standard components Figure 2 Accessories 2a 1 2b 4 5 A B D C 6 B A ...

Страница 5: ...n drain RS485 interface for service and extended field operation Modbus protocol Standard Firmware I1 48 2b Integrated Pump Driver IPD 600 2 50 01 100 10180 Housing ETFE chemical resistant coated alu waterproof IP65 1 Cable FEP jacket open wires cable length 5 m Table 2 Specification of standard components 1 Designed and tested for IP67 Pos Component Article Name Article Characteristics Value Feat...

Страница 6: ...oop flow or pressure control in connection with either a flow or pressure sensor see Figure 4 Precise ultrapure flow control systems can be realized with the pump system in combination with LEVIFLOW flowmeters Figure 4 Standard system configuration for process control Pressure or flow control with external sensor details in Section 4 1 1 Chemical Resistance Area Integrated Pump Driver Integrated P...

Страница 7: ...Storage humidity range relative humidity Extremes for transportation 15 95 non condensing Normal storage conditions Ambient temp 20 to 30 C Relative humidity 50 non condensing DC supply fluctuations 10 of nominal voltage Transient over voltages typically present on the mains supply Surge immunity according to EN 61000 4 5 tested together with certified AC DC power supply Pollution degree 2 Table 4...

Страница 8: ... 1 5 2 0 2 5 3 0 3 5 0 20 40 60 80 psi gallons min bar liters min 8000 rpm 9000 rpm 6000 rpm 7000 rpm Specific gravity 1 g cm3 Viscosity 0 7 cP Liquid Temp 40 C DC Voltage 48 VDC Pump Head LPP 600 29 5000 rpm 0 20 40 60 80 100 120 77 87 97 107 117 127 137 147 157 167 177 187 0 1 2 3 4 5 6 7 8 9 25 30 35 40 45 50 55 60 65 70 75 80 85 90 psi F bar Liquid Temperature in C Pump Head LPP 600 29 Testing...

Страница 9: ...bles Minimum Bending Radius Permanent Installation Minimum Bending Radius Sometimes Moved FEP 8 8 mm 7x Cable OD 15 x Cable OD PVC 9 5 mm 7x Cable OD 15 x Cable OD Table 5 Specifications for min bending radius of driver cable Note If not mentioned explicitly all the cables are not suited for constant dynamic bending and movement Tube 1 4pcs M6x25 Screws SS PTFE coated torque spec 50 Ncm For pump h...

Страница 10: ...g module WCM i600 1 in context with of driver IPD 600 x Note 1 Non tolerated dimensions are for reference only Note 2 For other configurations refer to according drawings Mounting Screws 4x M4x16mm Inox A4 PTFE coated Torque 60 cNM Remove 4 motor screws before mounting WCM i600 NPT 1 8 x 6 9 mm In outlet for water cooling fittings ...

Страница 11: ... Rare earth material 4 4 pcs M6 x 25 mm for pump motor mounting SS stainless steel PTFE coated screws Driver IPD 600 1 50 Epoxy PVC IPD 600 2 50 ETFE FEP 5 Sealing O ring and gasket for motor housing FPM FKM 6 Cable bushing Cable material PVDF housing with FPM sealing PVC or FEP cable jacket 7 Driver housing Epoxy or ETFE coating waterproof IP65 Coils and electromagnetic circuit potted with an epo...

Страница 12: ...s EN61000 6 2 and EN61000 6 4 using the specified standard power supply TSP 600 148M If power supplies other than the specified one is used or if the pump system is running on a 48 VDC bus with other devices such as sensors the use of a DC powerline filter may be necessary 3 3 Temperature Monitoring To avoid overheating of the system the controller and motor temperatures within the driver are moni...

Страница 13: ...for the IPD 600 2 50 driver 70 C liquid temperature Typical data for pumping with pump head LPP 600 29 temperature is measured inside of the integrated motor and controller contact temperature is below this temperature 77 97 117 137 157 177 0 0 5 0 10 0 15 0 20 0 25 35 45 55 65 75 85 0 20 40 60 80 F gallons min C liters min Ambient Temp 25 C Liquid Temp 25 C Specific gravity 1 g cm3 Viscosity 0 7 ...

Страница 14: ...thermal data are typical values which are partly based on measurements and partly on interpolations with a simplified thermal model and are therefore only guideline values and are suitable for a first layout of the basic thermal concept It is recommended to check the thermal values with the motor placed on the final location and under worst case performance conditions of the application To account...

Страница 15: ...e of the pump head and minimizes heat up of the liquid 6 In general the pump system placement and circuit shall be designed in a way that gas bubbles can leave the pump housing and that the pump head remains primed 7 To minimize heat up of the liquid the overall pressure drop in the hydraulic circuit shall be reduced as much as possible 8 It shall be avoided to pump for long periods of time agains...

Страница 16: ...t Voltage or current Output Actual Speed Actual Process Value 0 10 V 0 10000 rpm Speed Mode 0 10 V 0 100 Process Mode Direct connection no protection Galvanic isolation on user side is required AGND is reference The configuration of the analog output can be changed with Levitronix Service Software The analog output can be changed to a 4 20mA current output in the Levitronix Service Software Dout1 ...

Страница 17: ...tal Input 1 Blue Analog Input Ground Violet Analog Input 1 Grey Pink Analog Input 2 White Green Analog Ground GND Blue Red Analog Output Brown Green RS485 White Yellow RS485 Yellow Brown Do not connect White Grey Do not connect Cable Shield Relais GND 0 10 V 2 mA RS485 USB Cable with Termination Personal Computer Service 24 V 5 24 V 0 10 V R 5 kW For voltage output 0 10 V Analog inputs are not gal...

Страница 18: ...e pump system with a PLC the analog input can be used to set either the speed or the process value flow or pressure The digital and analog outputs can be used to monitor the pump status and operating parameters see Table 7 CAUTION The analog inputs and outputs are not galvanic isolated from the controller electronics To avoid ground loops and malfunction use floating analog signals 1 Power off the...

Страница 19: ...on 4 2 1 Mechanical Installation of Driver and Pump Head The driver can be fixed with four screws on the motor feet see Figure 7 Mounting can be done in either in horizontal or vertical position When mounting the pump head fittings into a circuit care shall be taken that no mechanical stress is acting on the fittings which can result in distorting creeping effects WARNING BPS i600 pump drivers are...

Страница 20: ... Pulse 300 700 ms ON Process Control Mode Status Active Error not active State 6 Internal Error Internal Error Enable not active Process Mode active Process Mode not active ON Priming Mode Status active Error not active State 10 Priming on Enable active Enable not active Timeout Process Mode not active Timeout Process Mode active Figure 17 PLC interface state diagram for standard firmware I1 48 Fo...

Страница 21: ...ording digital output on the PLC Interface Module is activated The pump system is switched OFF By a pulse off 300 700 ms on the Enable Reset input the system gets back to the Off state Errors Effect on Designated Digital Output of the PLC No pump head detected Error Open circuit Driver temperature was above 90 C Error Open circuit Driver temperature was higher than 80 C for more than 10 min Error ...

Страница 22: ...ash TOXIC or CORROSIVE CHEMICALS because of the strong magnetic forces Flush the pump housing before removing it WARNING HARMFUL CHEMICALS may be present Skin contact and toxic gases may be hazardous to your health Wear safety gloves and other appropriate safety equipment CAUTION Pay attention to the magnetic forces of the impeller when handling the pump head The attraction of magnetic parts and p...

Страница 23: ...ion with the pump system in connection with a PC and a USB interface The USB to RS485 adaptor cable as specified in Table 3 can be used to setup communication The software can be used for performing detailed troubleshooting The Levitronix Service Software is available in the download section on the Levitronix web page or contact the Levitronix Technical Service Department see under Section 8 8 Tec...

Страница 24: ...s for Fluids basic requirements are followed EN12162 Procedure for hydrostatic pressure testing in fluid pumps used for max pressure testing of pump head ISO12100 Safety for machinery principles for risk assessments used for system risk analysis EMC Directive 2014 30 EC The following standards of the EMC directive are tested and confirmed at a certified laboratory EN61000 6 2 Generic standards Imm...

Страница 25: ...not avoided could result in property damage Signal word SEMI S1 0701 Safety alert for Warning and Caution Safety alert SEMI S1 0701 Safety alert for Danger Safety alert SEMI S1 0701 Caution refer to accompanying documents is used on article labels for reference to manual Refer to manual ISO 3864 Toxic material poison Hazard identification IEC 61310 Corrosive material corrosion Hazard identificatio...