The dark field condenser is a system of lenses inserted into the cylinder. The cylinder is placed inside the microscope body on 

the moving ring. This ring is a centering device, which can be controlled with adjustment screws located on the outside of the 

microscope body.

The light from the illuminator, passing through the dark field condenser, is converted into a cone-shaped beam and directed at 

the observed specimen. Almost all of the light passes through the lens and forms an image only by the light scattered by the 

structure of the specimen. Therefore, in the field of view of the eyepiece, the bright contrast areas of the specimen structure 

are displayed on a dark background.
Assembly and using

 

• 

Install a dark field condenser instead of the standard condenser of your microscope.

• 

Install the desired eyepiece and choose the objective with the lowest magnification.

• 

Put a drop of oil on the condenser lens; fix the microscope slide on the stage. 

Note

: It is recommended to use a coverslip 

with a thickness of 0.8—1.2mm.

• 

Raise the dark field condenser until the immersion oil meets the stage and then focus.

• 

If you see an off-center light spot, use the adjustment screws to bring it to the center of the field of view.

•  Note

: The aperture of the objective lens used must be smaller than the aperture of the condenser.

Halogen bulb replacement

Unplug the microscope from a power source. Loosen the knurled screw on the underside of the microscope and open the panel 

to gain access to the lamp. Make sure the lamp has cooled down and then remove it. Do not touch the new lamp with your 

fingers; fingerprints or dirt will reduce the brightness and the lamp’s life. Use gloves to replace the lamp. If necessary, wipe the 

lamp with a clean, soft cloth. Use a lamp of a suitable type. Insert the contacts into the slot vertically. Then, close the panel 

and tighten the knurled screw finger tight.

Fuse replacement

Unplug the microscope from a power source. Open the fuse holder located on the back of the microscope body with a flat head 

screwdriver. Remove the old fuse and install a new one. Only use the appropriate types of fuses. Close the fuse holder.

Digital camera

Levenhuk MED D25T comes with a digital camera (fig. 2a), Levenhuk MED D25T LCD comes with a digital camera with an LCD 

screen (fig. 2b). A digital camera is installed to the third eyepiece tube and enables observing specimens on the screen in 

exquisite detail and true color, taking photos and videos, and saving the images for future reference.

Содержание MED 25

Страница 1: ...Levenhuk MED 25 Microscopes User Manual ...

Страница 2: ...htness adjustment knob 19 Binocular trinocular head locking screw 9 Aperture diaphragm adjustment rod 20 Third eyepiece tube trinocular models only 10 Coarse focusing knob 21 Tension adjusting ring 11 Fine focusing knob 2 1 3 2 1 Fig 1a Levenhuk MED 25B Fig 1b Levenhuk MED 25T D25T D25T LCD Fig 2a Levenhuk MED D25T 1 Digital camera 2 USB cable 3 Installation CD Fig 2b Levenhuk MED D25T LCD 1 Digit...

Страница 3: ...b until the specimen is about 3 175mm 1 8 away from the lens Looking into the eyepiece slowly turn the coarse focusing knob until you can see the image of the observed specimen Turn the fine focusing knob to make the image clear The fine focusing mechanism allows you to focus on the observed specimen while using high magnifications When you use an oil immersion objective the space between the spec...

Страница 4: ...ot use the adjustment screws to bring it to the center of the field of view Note The aperture of the objective lens used must be smaller than the aperture of the condenser Halogen bulb replacement Unplug the microscope from a power source Loosen the knurled screw on the underside of the microscope and open the panel to gain access to the lamp Make sure the lamp has cooled down and then remove it D...

Страница 5: ...double layer mechanical 140x140mm with a mechanical scale Focus system coaxial coarse 0 5mm and fine 0 002mm with rack and pinion mechanism coaxial coarse 0 5mm and fine 0 002mm with rack and pinion mechanism Condenser bright field Abbe N A 1 25 iris diaphragm and filter holder Abbe N A 1 25 iris diaphragm and filter holder Condenser dark field oil N A 1 36 1 25 with dovetail mount oil N A 1 36 1 ...

Страница 6: ...pening the base or changing the illumination lamp Regardless of the lamp type halogen or incandescent give it some time to cool down before trying to change it and always change it to a lamp of the same type Always use the power supply with the proper voltage i e indicated in the specifications of your new microscope Plugging the instrument into a different power outlet may damage the electric cir...

Отзывы: