background image

3

General Information

Read the user manual carefully before you start working with a microscope. Levenhuk MED Series biological microscopes are 

safe for health, life and property of the consumer and the environment when properly used, and meets the requirements of 

international standards. These microscopes are designed for observing transparent objects in the transmitted light using the 

bright field method and the dark field method. They perform well in clinical research and tests; teaching demonstrations; 

bacterioscopy and cytology in medical and health establishments, laboratories, universities; and may be used for scientific 

research in agriculture and microbiology.

Microscope assembly

• 

Unpack the microscope carefully and place it on a flat surface.

• 

Remove the plastic bags and dustproof cover of the eyepiece head.

• 

Install the binocular or trinocular head and tighten the locking screw.

• 

Inspect all the items included in the kit and define their purpose.

• 

Connect the power cord to the microscope and plug it in.

Microscope use: The bright-field method

Please refer to the fig. 1а (binocular model) and fig. 1b (trinocular model).

• 

Turn on the power and the illumination will turn on. Set the brightness at approximately 70%. Place the specimen on the 

stage. Make sure the clips hold the specimen firmly in place.

• 

Iris diaphragm should be adjusted in accordance with numerical aperture of a lens. The best practice is to make the iris 

diaphragm slightly smaller than the aperture of the selected lens. Open or close the iris diaphragm using the adjustment 

screws. If the adjustment screw is turned to the right, the iris diaphragm is fully open. 

Note

: the diaphragm is not intended 

for adjusting the brightness of the illumination. To adjust the brightness, use the brightness adjustment knob.

• 

Slide the filter holder, place the filter in it and then return the holder to its initial position.

• 

To change the objective magnification, turn the revolving nosepiece until it clicks.

• 

Make sure that the lens does not touch the sample when adjusting the focus: rotate the coarse focus knob until the 

specimen is about 3.175mm (1/8") away from the lens.

• 

Looking into the eyepiece, slowly turn the coarse focusing knob until you can see the image of the observed specimen. Turn 

the fine focusing knob to make the image clear. The fine focusing mechanism allows you to focus on the observed specimen 

while using high magnifications.

• 

When you use an oil immersion objective, the space between the specimen and the lens should be filled with immersion oil. 

Raise the condenser to the upper position and put a drop of oil on the lens and on the cover glass of the specimen. Move the 

revolving nosepiece back and forth to get rid of air bubbles in the oil. Then, fix the objective lens in its working position. 

Make sure that oil fills the entire space between the objective lens and the specimen. After use, wipe the lens dry.

• 

To adjust the position of the mechanical scale, turn the coarse focusing knob in a clockwise direction if you want to fix the 

position of the mechanical scale, or in a counterclockwise direction if you want to loosen it.

• 

To place the specimen in the field of view, move the stage horizontally back and forth or left and right using the stage 

moving knobs.

• 

Look through the right eyepiece with your right eye and adjust the image sharpness with the coarse and fine focusing knobs. 

Then, look through the left eyepiece with your left eye and rotate the diopter adjustment ring to equalize the difference 

between your left and right eyes vision. Remember the setting for future use. Adjust the distance between the eyepiece 

tubes of the binocular head so that the image merges into a single circle. 

• 

Trinocular models: you can install the digital camera to the third eyepiece tube. Use the beam splitter to send the light to 

the eyepieces or to the camera.

Dark field condenser (oil)

Configuration and operating principle

Please refer to fig 3. The device is designed for studying low-contrast objects that are not visible in the microscope when 

observed in normal light conditions using the bright field method. The dark field method allows you to observe very bright and 

transparent objects in scattered or reflected light. 

1

2

3

4

Fig. 3

1

Spherical mirror and cardioid lens, glued together and 

inserted into the frame

2

Cylinder

3 Body

4

Adjustment screws

Содержание MED 25

Страница 1: ...Levenhuk MED 25 Microscopes User Manual ...

Страница 2: ...htness adjustment knob 19 Binocular trinocular head locking screw 9 Aperture diaphragm adjustment rod 20 Third eyepiece tube trinocular models only 10 Coarse focusing knob 21 Tension adjusting ring 11 Fine focusing knob 2 1 3 2 1 Fig 1a Levenhuk MED 25B Fig 1b Levenhuk MED 25T D25T D25T LCD Fig 2a Levenhuk MED D25T 1 Digital camera 2 USB cable 3 Installation CD Fig 2b Levenhuk MED D25T LCD 1 Digit...

Страница 3: ...b until the specimen is about 3 175mm 1 8 away from the lens Looking into the eyepiece slowly turn the coarse focusing knob until you can see the image of the observed specimen Turn the fine focusing knob to make the image clear The fine focusing mechanism allows you to focus on the observed specimen while using high magnifications When you use an oil immersion objective the space between the spec...

Страница 4: ...ot use the adjustment screws to bring it to the center of the field of view Note The aperture of the objective lens used must be smaller than the aperture of the condenser Halogen bulb replacement Unplug the microscope from a power source Loosen the knurled screw on the underside of the microscope and open the panel to gain access to the lamp Make sure the lamp has cooled down and then remove it D...

Страница 5: ...double layer mechanical 140x140mm with a mechanical scale Focus system coaxial coarse 0 5mm and fine 0 002mm with rack and pinion mechanism coaxial coarse 0 5mm and fine 0 002mm with rack and pinion mechanism Condenser bright field Abbe N A 1 25 iris diaphragm and filter holder Abbe N A 1 25 iris diaphragm and filter holder Condenser dark field oil N A 1 36 1 25 with dovetail mount oil N A 1 36 1 ...

Страница 6: ...pening the base or changing the illumination lamp Regardless of the lamp type halogen or incandescent give it some time to cool down before trying to change it and always change it to a lamp of the same type Always use the power supply with the proper voltage i e indicated in the specifications of your new microscope Plugging the instrument into a different power outlet may damage the electric cir...

Отзывы: