LEM UNILAP 100 E Скачать руководство пользователя страница 3

Z

S

Measuring range

Resolution

Operating error

  man: 3 k

W

...300 M

W

1 k

W

 ...1 M

W

  

± (8% of m.v. + 1 digit)

auto: 3 k

W

 ... 10 M

W

1 k

W

 ...100 k

W

Measuring time:

As long as the "START“ button is pressed;
Subsequent automatic discharge of the
test piece via  400 k

W

Programmable limit:

RISO Limit:

1 k

W

 ... 299 M

W

 (man)

1 k

W

 ... 9.9 MW (auto)

AUTO:

Test sequence N - PE, L - PE, L - N,
programmable

Display of the actual measuring voltage:

Measuring range

Resolution

Operating error

1 ... 520 V DC

1 V

± (8 % of m.v. + 5 Digit)

Internal resistance:

Approx. 400 k

W

 (L/N-PE); serves as

discharge resistance for any capacitors in
the measuring circuit.

Max. interference
voltage:

Ueff = 

1

/

10

 of nominal output.

No measurement is started at higher
voltages.

Loop impedance (ZS/R) L-PE or L-N (L) IEC 61557-3
Measuring method:

Voltage drop

Nominal voltage:

95...145V, 175...300V,
330...440V (only L-N (L))

Reference  voltage:

110/230/400 V or 127/220/380 V AC or

measured voltage

Frequency range:

15.3 ... 17.5 Hz, 45 ... 65 Hz

Test current:

L-PE

L-N(L)

Voltage range

Test current

Zs = 0 

W

Zs = 200 

W

X

X

55...145 V

1.4...3.61 A

0.4...0.61 A

X

X

175...300 V

1.75...3.0 A

0.58...1 A

X

330...440 V

2.75...3.67 A

1...1.4 A

Measuring  range

Resolution

Operating error

0.07 ... 199 

W

0.01 

W

 ...1 

W

±  (5% of m.v. + 3 digit)

Measuring time:

Approx. 4 - 50 periods;
Duration as per IEC 1010

Mains imped. angle:

cos 

j

 > 0.5

Max.  inductance:

5 mH in voltage range > 175 V

Programmable limit:

Z

S LIMIT

: 0.01 ... 199 

W

Admiss overload:

Max. U

eff

 = 600 V AC

(measurement is not started outside the
admissible voltage and frequency ranges)

Short-circuit current

Range

Display

Resolution

Operating error

1 A…10 kA

1…40kA

1…10…100 A

 results from I

=U

N

U

N

 as selected:

1: 110V, 230V, 400V. 2: 127V, 220V, 380V
resp.  3: measured voltage

Valid ranges for U

N

:

95…145V, 175…300V, 330…440V
(only at  Z

S L-N

)

Valid ranges for frequency: 15.3Hz…17.5Hz, 45Hz … 65Hz

Earthelectrode fault voltage (SEV 3569)
Earthelectrode voltage, with probe only

Range

Resolution

Measuring method

0…U

N

0.1 V

U

F

 =  R

. I

K

Detection of rotary field direction IEC 61557-7
Voltage range:

20 ... 440 V AC, 15.3 ... 65 Hz

Admiss. overload:

Max. U

eff

 = 600 V AC

Max. current to earth:

< 3.5 mA

Rotary direction display: Symbol for right- / left-handed
Internal resistance:

200 k

W

 ... 400 k

W

"Elliptic rotary fields" with two L-conductors and the neutral conduc-
tor can also be tested.

Low resistance measurements (R1k

W

) IEC 61557-4

Measuring method:

Current / voltage measurement  with
automatic pole reversal.

Open-circuit  voltage:

Approx. 20 V DC

Short-circuit current:

> 200 mA DC

Admiss. overload:

Max. U

eff

 = 600 V (before START),

no START for > 5 V

Measuring  range

Resolution

Operating error

0.12 ... 2.99...19.9 

W

...1 k

W

0.01...0.1...1 

W

± (5 % of m.v. + 3 digits)

Measuring time:

Approx. 2 s incl. voltage polarity reversal.

Continous measurement:With depressed START button
Admissible  inductance:

Max. 5 H

Programmable limit:

R

LIMIT

  = 0.01 

W

 ... 999 

W

Series-mode
interference  voltage:

Max. 40 V

eff

 AC, in case of higher  voltage

measurement is terminated.

Measuring protocol-print in format A4:

WinSAT 100

the complete system for measured data archiving and protocol tool
for professional electricians.
Application software for UNILAP 100 E and UNILAP 100 XE.
Multilingual, integrated project management and data export.
Can run under Microsoft Windows® 98 and Windows® NT.

Functions

Code-programming of the UNILAP 100 (X)E
Using tabs, WINSAT100 provides for easy and clear set-up of all
limit values and settings stored inside the device, as well as the
device’s individual configuration with regard to any application
purpose, and to applicable regional regulations.

Installation descriptions in a tree structure with additional
information can be printed out together with bar codes and used as
a check plan. Same manipulation as Windows

®

 explorer,