Technical Data

Display:

3-

1

/2 digit (1999 ), 7-Segment-Liquid-Crys-

tal-Display, 17 mm high, with illumination

Operating temperature
range:

0° C … + 35° C

Working temperature
range:

- 10° C … + 50° C

Storage temperature
range:

- 20° C … + 60° C

Reference temperature
range:

23° C ±2° C

Intrinsic error:

Refers to the reference temperature

Temperature coefficient: ± 0.1 % of m.v. / K
Operating error:

Refers to the operating temperature range

IEC 61557-1

Climatic class:

JWG as per DIN 40040 and IEC 654-1;
Relative humidity 65 % annual mean, 85 %
max., no damp.

Protection:

IP 40 as per DIN 40050 and IEC 529-2

Safety class:

class II (      ) 300 V, install. category III
as per IEC 1010-1 / EN 61010-1,
pollution degree 2.

Test  voltage:

3700 V as per IEC 1010-1 / EN 61010-1

Input protection:

By software and varistors against voltages
U

eff

 > 600 V and quick-acting fuses

(6.3 A / 500 V)

Max. voltage to earth:

Ueff = 300 V

Clearance / creepage
distances:

Correspond  to IEC 1010-1 / EN 61010-1

Emission:

Class B as per EN 50081-1 and IEC 61326-1

Immunity:

Class A as per EN 50082-1 and IEC 61326-1

Quality standard:

as per DIN ISO 9001

Auxiliary  power:

6 pcs. 1.5 V alkaline manganese batteries
(IEC LR 6) or accu 7.2V / 1500 mAh
(option).
Internal battery for date / time - life 5
years.

Dimensions:

265 x 265 x 90 mm (L x W x H) incl. lid
and compartment for accessories

Weight:

approx. 2.3 kg w/o batteries and w/o
accessories
approx. 5.7 kg in carrying case

IrDA®  Interface (Infrared Data Association) as standard. For easy
PC-communication (remote control, data acquisition, reading of
stored data). RS 232 interface possible as option

.

Integrated data memory for  255 data records (approx. 6000
measurement values).
Integrated real time clock with date. Barcode-reader for up to 18
characters useable (option RS 232 interface necessary).
The limits prescribed  by the standards can be adapted individually.
A limit infringement  is indicated by audible and optical warnings.
Automatic compensation for  standard accessories.
Compensation for line extensions up to 5 

W

 possible.

Protective conductor check
Nominal voltage:

50 V … 300 V AC  / 15.3 ... 100 Hz
between contact electrode and  PE-line

Internal  resistance:

Approx. 1.5 M

W

Phase indication
Nominal voltage:

20 V … 300 V AC / 15.3…420 Hz

Internal  resistance:

Approx.  400 k

W

Admiss. overload:

Max. U

eff

 = 600 V

Note:

The voltage of the "L“ and "N“ wire is
measured to "PE“, evaluated and a
symbol indicates the live contact.

Voltage (DC/ AC), Frequency

Range

Resolution

Frequency Range

Operating error

0..50..440…550V

1 V

DC, 15.3…420 Hz

 ±(1% of mv +1dig)

15.3..99.9..420Hz

0.1…1Hz

5…440 V

 ±(0.1% of mv +1dig)

Internal resistance:

300…400 k

W

 (L – N – PE)

Residual current operated device check
(FI-RCD / IEC 61557-6)
Measuring functions:

RCD-test without tripping

Tripping test with pulses or ramp
(tripping time, tripping current)
Fault voltage
Loop impedance, short-circuit current
(without tripping)
Earth resistance (with probe)
Standard/ selective RCDs

Test currents:

I

D

N

 x1, x2, x5

positive, negative phase position
positive, negative pulsating DC current
pulses, current steps

Voltage range:

95 ... 145 V, 175 ... 300 V
Frequency range: 15.3...17.5 Hz, 45...65 Hz

Admissible  overload:

Max. Ueff = 600 V

Function

 

Intrinsic error of

Remark*)

test current

Ramp

+/- 10 %

18 steps, 28...125 of I

D

N

Pulse,

tripping test

0...+10 %

x1/x2/x5 I

D

N

, 150 mA, 250 mA

Pulse,

non tripping test

-10 ... 0 %

0.3 (0.5) I

D

N

*) Rated residual current I

D

N

 = 10, 30, 100, 300, 500 or

var 6 ... 1000 mA. For function "halfwave/DC" and / or x1/x2/x5
there are some restrictions due to the increased peak current for
higher rated residual currents.

 Fault voltage range (U

F

)

Resolution

Operating error

0.5 … 99.9 V

0.1 V

0…± 8 % of mv. + 2 digit)

Automatic test stop: U

F

 

>50 V complies with IEC 1010

Tripping time (t

A

)

Resolution

Operating error

0 … 500 ms (300 ms)

   1 ms

     ± 2 ms

Loop-

Resolution

 Operating error

impedance Z

S

(

W

)

resp. earth resistance R

A

0.2  

W

…9.99  k

W

0.01 

W

…10 

W

± (10% of mv. + 4 digit)

Test current period as per IEC 1010-1. Limitation of duration
period taking into account the fault voltage as per IEC 61557-6
and IEC 1010-1.

Positive or negative pulsating direct current:

Tripping test - corresponding to the applicable standards which
admit 0.35 ... 1.4 I

D

N

 as tripping current. For ramp function the

tripping current is displayed

 

as TRMS of  the half-wave current.

Earthing resistance (RA) IEC 61557-5
Measuring method:

Current / voltage measurement with probe

Voltage  ranges:

95 ... 145 V, 175 ... 300 V, outside
these ranges will not be started.

Frequency ranges:

15.3 ... 17.5 Hz, 45 ... 65 Hz

Admiss overload:

Max. U

eff

 = 600 V before start,

(for > 5 V no start), termination of
measurement for U

eff

 > 50 V

Measuring time:

max. as per IEC 1010, 2 ... 26 periods

Automatic compensation for  standard accessories.
Compensation for line extensions up to 5 

W

 possible.

Probe voltage

Resolution

Operating error

1 ... 70 V

1 V

± (2 % of m.v. + 1 digit)

Measuring range

Resolution

Operating error

0.01 

W

...0,15 

W

...10 k

W

0.1 

W

 ... 10 

W

± (10% of m.v. + 3 digit)

Test current:

1 A for < 20 

W

Max. interference voltage: When U

S-PE

 > 20V no measurement

Max. probe resistance:

10 k

W

, for (RA + Rprobe) > 20 k

W

 no start

Programmable limits:

0.01 

W

 ... 9.99 k

W

Insulation resistance (RISO) IEC 61557-2
Measuring method:

Current / voltage measurement

Nominal output voltage:

100 / 250 / 500 V DC

Open-circuit  voltage:

Approx. 105 / 260 / 520 V DC

Nominal  current:

> 1 mA DC (>2.5 mA DC at 250 V)

Short-circuit  current:

< 7 mA DC

Admiss. overload:

Max. U

eff

 = 600 V AC; (test is locked)