140528/2
Page
21
of
32
5.3
Grounding the transducer head
For safety reasons all types of transducer heads must be properly grounded.
Follow the description below for the appropriate transducer head in order to
establish a good grounding connection.
ITZ 600-25-PR
: A Ground wire (min 1.5 mm
2
) with an Ø 4.3 mm ring tongue
must be connected to the transducer head. The ring tongue is fastened to the
M4 stud in the lower rightmost corner of the STH600 using a M4 nut.
ITZ 2000-50 (-PR)
: A ground wire (min 1.5 mm
2
) with an Ø 8.3 mm ring
tongue must be connected to the transducer head. The ring tongue is placed
on one of the 4 mounting screws before mounting the head.
ITZ 5000-140 and ITZ 10000-100
: A Ground wire (min 1.5 mm
2
) with an Ø
10.8 mm ring tongue must be mounted to the transducer head. The ring
tongue is placed on one of the 4 mounting screws before mounting the head.
ITZ 16000-150 and ITZ 24000-150
: A Ground wire (min 1.5 mm
2
) with an Ø
12.3 mm ring tongue must be mounted to the transducer head. The ring
tongue is placed on one of the 2 mounting screws before mounting the head.
5.4
Installation
1. Establish the ground connection (see section 5.3).
2. Mount the provided connection cable between the ITZ FLEX ULTRASTAB
electronics and transducer head.
3. Connect the Analog Output Cable as described in section 13.4 and 13.5.
4. Check that all cables terminated in a plug are correctly installed and that
the lock screws are tightened firmly.
5. Connect the supplied power cord to the IEC inlet on the unit to turn it on.
The transducer head and electronics can be installed with a transducer head
connecting cable of 30 meters maximum.