General
6
Analyst
3P
EO 0620GDF Rev. E
1.2
Safety Instructions
Please read this section carefully. It makes you familiar with the most
important safety instructions for handling your device.
Important
This power quality analyser may only be used and handled
by qualified personnel.
Protect the device against wetness and humidity, so that
the device does not suffer damage.
The plug and socket connection for the voltage input is
designed for 600V CAT III. The maximum voltage between
outer conductor and earth potential must not exceed 600V.
With multi-phase connections, the voltage between the outer
conductors of the system to be measured may not exceed
800V.
Warning
Current probes must comply to IEC 61010-2-032 category A
and must be designed for 600V CAT III. If flexible current
probes are used, the conductor to be enclosed by the probe
and the adjacent conductor must be connected strain-free,
or suitable protective gloves must be worn.
Warning
Open the device only to replace the accumulator package
(see chapter 4.1). Opening the device can lead to electrical
shocks. Therefore the device has to be separated from all
live sections before opening. Also disconnect all test leads
before using the RS232 interface. Maintenance work can
only be done by qualified service personnel.