21
Fasten test piece
• Release tension
clamping jaws (11)
with the
clamping lever (9)
.
• Adjust
clamping jaws (11)
at the
adjusting screw (10)
to thickness of sample.
• Fix sample material with
clamping lever (9)
.
• Do not operate the sliding carriage backwards when the sample is
fastened in, overload fuse will be activated.
Shear test
Peel test
Tensile test
Start test procedure
• Press Start/Stop
›
key
• When the set initial tension value is achieved, the elongation and position values are reset to nought and the test
evaluation starts (if the initial tension value is already set to
O N / 0 lbf
, the test evaluation starts immediately).
•
During the test the parameters cannot be changed.
End test procedure
• In case the sample tears, push key
›
to stop
sliding carriage (12).
• If key
›
is not pressed, the
sliding carriage (12)
stops automatically at the end.
• To interrupt or stop the tensile test, press
›
Start/Stop key.
• When sample shears, the
sliding carriage (12)
stops automatically.
• If the sample does not shear, the
sliding carriage (12)
stops at the end of the run.
• To interrupt or stop the tensile test, press
›
Start/Stop key. When interrupting the tensile test, the measu-
ring values are not reset if the power sensor is under pressure the initial tension value (see F
Peak
display).
This guarantees that the existing test can be continued.
• The automatic shut-off can be a hindrance during certain tensile tests when the sample tears (FTear).
Therefore, this function can be deactivated.
Go to the selection menu by pressing the
›
and
–
‹‹
keys
at the same time.
With the
+
››
key the automatic shut-off is active (default) and
with
–
‹‹
key it is deactivated (manual).
–
‹‹
<
< - > M a n u a l - S t o p
+> A u t o - S t o p
&
›