Guidance and manufacturer’s declaration - electromagnetic immunity

Test 

Frequency

 (MHz)

385

0.3

27

The device is intended for use in the electromagnetic environment specified below. The customer or the 

user of the device, should assure that it is used in such an environment.

NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna 

and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by 

IEC 61000-4-3.

Radiated RF 

IEC61000-4-3 

(Test 

specifications 

for 

ENCLOSURE 

PORT 

IMMUNITY to 

RF wireless 

communicatio

ns equipment) 

Band a)

(MHz)

Service a) Modulation b) Modulation b)

(W)

Distance (m) IMMUNITY 

TEST 

LEVEL 

(V/m)

380-390

TETRA 

400

Pulse 

modulation b) 

18Hz

1.8

450

430-470

GMRS 460

FRS 460

FM c) ± 5kHz

deviation 1kHz

sine

2

0.3

28

710

704-787

745
780

LTE Band

13,

17

Pulse 

modulation b)

217Hz

0.2

0.3

9

810

870

930

800-960

GSM 

800/900,

TETRA 800,

iDEN 820,

CDMA 850,

LTE Band 5

Pulse 

modulation b)

18Hz

2

0.3

28

1720

1845

1970

1700-

1990

GSM 1800;

CDMA 1900;

GSM 1900;

DECT;

LTE Band 1,

3,

4,25; UMTS

Pulse 

modulation b)

217Hz

2

0.3

28

2

0.3

28

2450

2400-

2570

Bluetooth,

WLAN,

802.11 

b/g/n, RFID

2450, LTE

Band 7

Pulse 

modulation b)

217 Hz

5240

5500

5785

5100-

5800

WLAN

802.11

a/n

Pulse

modulation b)

217 Hz

0.2

0.3

9

a) For some services, only the uplink frequencies are included.

b) The carrier shall be modulated using a 50% duty cycle square wave signal.

c) As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because while it does 

not represent actual modulation, it would be worst case.

The MANUFACTURER should consider reducing the minimum separation distance, based on RISK 

MANAGEMENT, and using higher IMMUNITY TEST LEVELS that are appropriate for the reduced 

minimum separation distance. Minimum separation distances for higher IMMUNITY TEST LEVELS shall be 

calculated using the following equation:

                                                                E=

Where P is the maximum power in W, d is the minimum separation distance in m, and E is the IMMUNITY 

TEST LEVEL in V/m.

Содержание LS810-BS

Страница 1: ...钰 2019 04 29 谭文昊 2019 04 29 LS810 BS 004 GB 06 PO 2B4069 LS810 BS 004 GB 06 EN说明书 A3 共 32 张 第 1 张 A 3 对应结构图纸 印色 1 1 表面处理 EC REP MDSS Medical Device Safety Service GmbH Schiffgraben 41 30175 Hannover Germany Guangdong Transtek Medical Electronics Co Ltd Zone A No 105 Dongli Road Torch Development District Zhongshan 528437 Guangdong China ...

Страница 2: ......

Страница 3: ......

Страница 4: ... 93 Power supply 3 7V 420mAH Built in rechargeable lithium ion battery 5V 1A USB ...

Страница 5: ......

Страница 6: ......

Страница 7: ......

Страница 8: ......

Страница 9: ......

Страница 10: ......

Страница 11: ......

Страница 12: ......

Страница 13: ......

Страница 14: ......

Страница 15: ......

Страница 16: ......

Страница 17: ......

Страница 18: ......

Страница 19: ......

Страница 20: ......

Страница 21: ......

Страница 22: ......

Страница 23: ......

Страница 24: ......

Страница 25: ......

Страница 26: ... other devices that emit electromagnetic fields near the unit This may result in incorrect operation of the unit 3 Caution This unit has been thoroughly tested and inspected to assure proper performance and operation 4 Caution This machine should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary this machine should be observed to verify normal...

Страница 27: ......

Страница 28: ...d 0 35 d 1 2 p 1 2 2 3 ...

Страница 29: ...1 10 100 12 12 23 2 3 ...

Страница 30: ... 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation b 18Hz 2 0 3 28 1720 1845 1970 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation b 217Hz 2 0 3 28 2 0 3 28 2450 2400 2570 Bluetooth WLAN 802 11 b g n RFID 2450 LTE Band 7 Pulse modulation b 217 Hz 5240 5500 5785 5100 5800 WLAN 802 11 a n Pulse modulation b 217 Hz 0 2 0 3 9 a For some services only the upl...

Страница 31: ......

Страница 32: ......

Отзывы: