When the second measurement has been taken, the
accuracy of ATR1 and, if previously selected, the
accuracy of the index and the line-of-sight errors are
displayed.
Repeats the ATR1 error determination procedure.
The calibration process is interrupted. The old values
will be left intact.
No further repeat measurements are required. The old
and the newly-defined ATR1 collimation errors become
optional, and are displayed together with the
line-of-sight errors (c) and the V-index errors (i).
The calibration can be repeated as often as necessary
until the required level of accuracy is obtained. The
result is the mean of all the measurements taken. It is
recommended that at least 2 measuring sequences be
carried out.
CAL\ ATR ACCURACY
14:03
No. of Meas :
2
σ
ATR Hz
:
0°00'05"
σ
ATR V
:
-0°00'08"
σ
V-Index
:
-----
σ
Hz-col
:
-----
Perform further measurements?
ABORT
NO
YES
F1
F2
F3
F4
F5
F6
HELP
CAL\ATR COLLIM. ERRORS
14:03
old new
ATR Hz-c :
0°00'08"
0°00'05"
ATR V-indx:
0°00'10"
0°00'09"
i V-indx :
0°00'00"
0°00'10"
c Hz-col :
0°00'10"
0°00'02"
Accept new value(s)?
RETRY
NO
YES
F1
F2
F3
F4
F5
F6
HELP