14
By using the depletion laser 592nm, the samples can be bleach very quickly, so first scan the sample in
standard
Confocal
mode to find the focus and prepare the STED configuration close to work.
Set the WLL laser line that you need to use (
10 to 30 %
).
Select the Hyd 2 detector and set it at
100%.
Introduce to the Hyd a bandwidth that is
10nm fahter
to the WLL line that you set, and maximum
10nm lower
than the wavelength of the laser depletion ( +/-582nm).
Use a format like
512x512
or
1024x1024
.
Use a scan speed like
400HRZ
.
Use a zoom factor from
4 to 6
.
STED configuration ( step1)