Leica S Series 

User Manual 

9

Safety Instructions (continued)

Legal regulations

 

Observe the generally applicable statu-
tory and country-specific regulations for 
accident prevention and environmental 
protection.

EC Declaration of Conformity

 

Electrically operated accessories are 
constructed based on the state of the art 
of technology and are provided with an EC 
Declaration of Conformity. See page 50.

Health risks

Workplaces with stereomicroscopes facilitate 
and improve the viewing task, but they also 
impose high demands on the eyes and holding 
muscles of the user. Depending on the dura-
tion of uninterrupted work, asthenopia and 
musculoskeletal problems may occur. For this 
reason, appropriate measures for reduction of 
the workload must be taken:
Optimal arrangement of workplace, work 
assignments and work flow (changing tasks 
frequently). Thorough training of the person-
nel, giving consideration to ergonomic and 
organizational aspects.
The ergonomic design and construction of the 
Leica M stereomicroscopy series are intended to 
reduce the exertion of the user to a minimum.

Direct contact with eyepieces can be a potential 
transmission path for bacterial and viral infec-
tions of the eye.

The risk can be kept to a minimum by using 
personal eyepieces for each individual or 
detachable eyecups.

Содержание S8 APO B

Страница 1: ...Leica S Series User Manual 1 Living up to Life Leica S8 APO B User Manual...

Страница 2: ...ons in whole or in part by print photocopy microfilm or other method including electronic systems is not allowed without express written permission from Leica Microsystems CMS GmbH The instructions co...

Страница 3: ...cturer assumes no liabil ity for damage caused by or any risks arising from using the microscope for other purposes than those for which they are intended or not using them within the speci fications...

Страница 4: ...ffer other accesso ries for use in clean rooms The cleaning of glass surfaces and objec tives in particular should be carried out exclusively as outlined in the brochure Cleaning of Microscope Optics...

Страница 5: ...es etc Please read the User Manual and the safety instructions from the supplier Before installing operating or using the instru ments read the user manuals listed above In particular please follow al...

Страница 6: ...uments Danger due to hot surface This symbol warns against touching hot surfaces e g those of light bulbs Important information This symbol indicates additional infor mation or explanations that are i...

Страница 7: ...ts and accessories described in thisUserManualhavebeentestedforsafetyand potential hazards The responsible Leica affiliate must be consulted whenever the instrument is altered modified or used in conj...

Страница 8: ...e original packaging for shipping or transporting the individual modules of the Leica M stereomicroscopy series and the accessory components In order to prevent damage from vibrations disassemble all...

Страница 9: ...uscles of the user Depending on the dura tion of uninterrupted work asthenopia and musculoskeletal problems may occur For this reason appropriate measures for reduction of the workload must be taken O...

Страница 10: ...n setting up the instruments If it is specified that two or more people are required for setup compli ance with this is mandatory Never spill any liquids on electrical instru ments This could cause th...

Страница 11: ...the Leica S8 APO B 25 Tips for Working Ergonomically 26 Using the Eyepieces 27 The Correct Interpupillary Distance 28 Focusing 29 Changing Magnification Zooming 30 Limiting Zoom Range 31 Limiting Zoom...

Страница 12: ...s User Manual 12 Technical Data Technical Data 45 Leica TL ST Transmitted Light Base 46 Appendix Calculating the Total Magnification Field of View Diameter 48 Troubleshooting 49 Care Maintenance Conta...

Страница 13: ...Leica S Series User Manual 13 Leica S8 APO B...

Страница 14: ...n a precise sharp display of your microscopic specimens at both large working distances and large fields of view from the complete image down to the finest enlarged detail Though the reliability and r...

Страница 15: ...m corrected center of the objec tive is used for the image This provides supe rior optical performance with large level and undistorted fields of view and chromatically optimized high contrast images...

Страница 16: ...assembled and commissioned by your Leica consultant click here to skip through the installation instructions and go directly to the Quick Start Guide on page 24 If on the other hand you are assemblin...

Страница 17: ...Leica S Series User Manual 17 Assembly...

Страница 18: ...p surface Refer also to the separately provided User Manual for the TL ST transmitted light base Focusing drive and column Never unscrew the 3 screws on the right side of the focusing column 1 Unscrew...

Страница 19: ...t in the desired position with the clamping screw Additional objective optional 1 Screw the desired objective counterclock wise into the optics carrier Protective objective glass optional 1 Screwtheob...

Страница 20: ...cule before you set the eyepiece in place Available graticules The following graticules and stage micrometers for calibrating may be ordered Graticule 10 mm 0 1 mm Graticule 5 mm 0 1 mm Graticule 5 mm...

Страница 21: ...d The scale must not appear reversed 2 Remove the insert from the bottom of the eyepiece and place it on the bench with the knurled side down 3 Hold the graticule by the edges to avoid leaving fingerp...

Страница 22: ...StereoZoom Leica S8 APO B with two adjustable eyepieces Inserting the eyepieces 1 Push the eyepieces as far as they will go into the tubes 2 Check that the eyepieces are seated firmly and precisely in...

Страница 23: ...ive for more details Installing the camera 1 Remove the protective dust cover from the video projection lens C mount adapter and the microscope camera 2 Screw the camera into the video projection lens...

Страница 24: ...Leica S Series User Manual 24 Quick Start Guide...

Страница 25: ...fication changer right drive knob with magnification scale 4 Fixing screw for optics carrier in the microscope carrier 5 TL3000 ST transmitted light base 6 Focusing drive 7 Stop for zoom limit 8 C mou...

Страница 26: ...take full advantage of its outstanding optical and ergonomic advantages Align your workstation optimally Consider the height of the bench and chair Use the whole seat surface and the back rest Ensure...

Страница 27: ...er the parfocality requires a personal dioptric correction for the user Dioptric correction In order to parfocally match the stereomi croscope at least one eyepiece with dioptric correction is necessa...

Страница 28: ...become automatic Reference values The interpupillary distance can be set between 55 and 75 mm An exit pupil distance is the distance between eye and eyepiece With the 10 23B wide field eyepiece for e...

Страница 29: ...tive The focusing drive can be operated either left or right handed Focusing 1 Align the specimen under the objective 2 Set the lowest magnification level At the lowest magnification the desired speci...

Страница 30: ...can be operated with the left and the right hand The image scale is shown on the right drive knob The basis for the calculation of the total magnification and the field of view can be found on page 48...

Страница 31: ...he following exam ple shows the limit to the range between 1 to 3 2 Defining the lower limit 1 Loosen the Allen screws on the left drive knob with the Allen key provided 2 Turn the right drive knob to...

Страница 32: ...rive knob until it touches the built in zoom stop 4 Carefully tighten the Allen screw Defining the upper limit stop 1 Loosen the Allen screws on the right drive knob with the Allen key provided 2 Turn...

Страница 33: ...end to slide downwards The resistance can be adjusted individually depending on the equipment weight and personal preferences as follows 1 Grip the outer drive knobs with both hands and turn them towa...

Страница 34: ...ide ways in the microscope carrier if the user wants to work from the side Changing position 1 Unscrew the clamping screw 2 Turn the optics carrier laterally to the desired position 3 Carefully tighte...

Страница 35: ...d only if you want to view a specimen detail that is located higher or lower Use this advantage as often as possible it is not available on all stereomicroscopes The diopters can be set between 5 and...

Страница 36: ...in the direction until this eye sees the specimen sharply 12 Repeat steps 10 and 11 with the other eye Continued on next page 6 Set the highest magnification level 7 Optimize the focusing with the foc...

Страница 37: ...nification level 2 Observe the specimen if necessary refocus it slightly 3 Change from the highest to the lowest magnification The sharpness should be constant parfocal If this is not the case repeat...

Страница 38: ...Controller for light intensity 4 Adjustment for path folding mirror Rear side of the transmitted light base TL ST 1 Screws for changing the halogen lamp 2 Power connection socket 3 Main power switch...

Страница 39: ...ved The mirror is kept in the correct position at all times and permits smooth changeover between bright field and opaque transmitted light Bright field Bright field is suitable for examining translu...

Страница 40: ...s with your bare fingers this drastically reduces the service life of the lamp Changing lamps 1 Unscrew the two screws on the cooling unit and pull the cooling unit out along with the lamp 2 Carefully...

Страница 41: ...ing and the framing are done with the left eyepiece video photo beam path Flat specimens are sometimes blurry on the left and right edge of the image This blurring is based on laws of optics and does...

Страница 42: ...Leica S Series User Manual 42 Dimensional Drawings in mm...

Страница 43: ...ted Light Illumination 207 107 25 116 75 351 3 8 22 278 58 76 144 104 113 148 205 278 205 171 207 107 25 116 75 351 3 8 22 65 223 278 58 76 144 104 113 148 205 220 278 205 171 207 107 25 116 75 351 3...

Страница 44: ...Leica S Series User Manual 44 Technical Data...

Страница 45: ...100 V Magnification basic configuration 10 80 Maximum resolution 600 lp mm Maximum numerical aperture 0 2 Working distance basic config 75 mm Object field diameter 23 mm Adjustable zoom limits 2 Vide...

Страница 46: ...ent temperature 10 40 C Connections Power plug Weight 7 4 kg Illumination modes Bright field Yes Dark field Yes single sided Oblique light No Relief Contrast System RC No CCIC Constant Color Intensity...

Страница 47: ...Leica S Series User Manual 47 Appendix...

Страница 48: ...ification changer position NFOV Field number of the eyepiece Field numbers are printed on the eyepieces 10 23 16 16 20 12 10 23B 16 14B 25 9 5B 40 6B Example MO Additional objective 1 6 ME 20 12 eyepi...

Страница 49: ...ed devices always first check Is the voltage selector set correctly Is the main power switch switched on Is the power cable connected correctly Are all connecting cables attached correctly Are the fus...

Страница 50: ...ing from careless or improper handling EC Declaration of Conformity To download the EC Declaration of Conformity use this link http www leica microsystems com products stereo microscopes macroscopes r...

Страница 51: ...onents Some components are made of polymer or are polymer coated They are therefore pleasant and convenient to handle The use of unsuit able cleaning agents and techniques can damage polymers Permitte...

Страница 52: ...ms CMS GmbH Ernst Leitz Stra e 35578 Wetzlar Germany 2013 Phone 49 6441 29 0 Fax 49 6441 29 2599 LEICA and the Leica logos are registered trademarks of Leica IR GmbH Printed on chlorine free bleached...