Control unit with touch panel
48
PROVIDO Ref. 10732418 / Version 03
PROVIDO Ref. 10732418 / Version 03 49
9.4 Menu – Maintenance menu
X
Press the Menu button and select "Maintenance".
The Maintenance menu offers the following screens:
• Lamp History
• Check Switches
• Microscope Settings
• Versions
9.4.1 Maintenance –> Lamp History
On this screen, you can view and reset the operating hours of
lamp 1 and lamp 2.
Whenever you replace a bulb, reset the bulb's hour meter
to 0 by double-clicking the "Reset" button.
A dialog window informs you when the xenon lamp is losing
luminosity and is no longer sufficient.
9.4.2 Maintenance –> Check Switches
On this screen, you can test your handles and the optional
footswitch.
Top right selection field
In this field you can select the connection you are using or the
desired handle.
X
Scroll forwards or backwards in the list by clicking the
arrowheads to select the connection.
Bottom right selection field
In this field you can select the footswitch you want to check.
X
Scroll forwards or backwards in the list by clicking the
arrowheads to select the footswitch.
X
Press all of the keys, one after the other, of the footswitch or
handle you want to test.
If the key you have pressed is functioning properly, a green dot
appears on it on the display. The comment "Tested" appears in
the caption field of the key.