5
Leica MATS – Safety concept
The Leica thermocontrol system MATS may only be used in
closed facilities. The following installation sites must be
excluded:
• Where flammable gas, corrosive gas oil mist and particles
that can deteriorate electrical insulation are generated or are
abundant.
• Where highly intense vibration or impact is generated or
transferred.
• Near high voltage power lines or where inductive interfer-
ence can affect the operation of the Leica thermocontrol
system MATS.
• Sites which create condensation water or humidity or are
exposed to direct sunlight.
The process value displayed at the control unit corresponds to
the temperature of the table surface. If the specimen in a petri
dish or on a slide glass indicates a lower temperature
compared to the displayed process value, adjust the set value
to obtain a proper specimen temperature. Otherwise, the
specimen could be damaged.
If the sensor malfunctions, the surface may become
very hot.
•
Dielectric strength: 2 sec. at 1350 V AC between power
supply terminal and ground terminal
•
Insulation resistance: 500 V Mega more than 100 M
Ω
minimum between power supply connection and ground
connection
•
Fire protection: Temperature cannot be set higher than 50 °C
•
Fuse: T AC250V 1.6 A
•
The Leica thermocontrol system MATS is electromagneti-
cally compatible in accordance with the directive 89/336/EEC
and can be operated at the same time as other electrical
appliances.
Place of use
Temperature control
Protective measures
of the manufacturer
Safety regulations
Содержание MATS
Страница 1: ...Leica MATS User manual...
Страница 9: ...9 Leica MATS Assembly MATS MZ367 LT MATS LZ LT MATS URS LT MATS UDM LT MATS U55S LT MATS STD LT...
Страница 18: ...18 Leica MATS User Manual Notes...
Страница 19: ...19 Leica MATS User Manual Notes...