19
HistoCore Water Bath
Setting up the instrument
4
2. Use the instrument with the provided power cable only. Insert the connector of the power cable
into the connection socket
and plug the power plug into the power socket.
Fig. 16
1
4.5 Switching on/off the instrument
Warning
The user plugs the instrument onto a power board with other instruments
.
The instrument malfunctions due to unstable current/power, which leads to tissue damage.
•
Do not use a power board to install the power cable.
•
The instrument must be connected to a grounded power socket.
Switching on the instrument
1.
Before switching the instrument on, fill the water bath with a sufficient quantity of distilled water.
2. Switch on the instrument with the main switch at the front right corner of the instrument. The
run/stop button on the control panel is illuminated.
3. Touch the run/stop button, the heating to the water bath and slide dryer(s) (if connected) starts.
For the buttons' functions on the control panel, go to
p. 21 – 5.1 Control panel on the instrument)
Switching off the instrument
The instrument is switched off by single touch of the run/stop button.
You do not have to switch the main switch on or off each time for everyday routine use.
4.6 Moving the instrument
The instrument must be switched off and disconnected from the power supply. The water tray must be
empty. The instrument must have cooled down before moving.