60
9. Contrast Methods
Special darkfi eld condensers are available for
the Leica DM2700 M (Fig. 61).
The application potential of the DF condensers
depends on the aperture of the objective in use.
For objectives with a built-in iris diaphragm, the
aperture can be adapted..
DF condenser
max. objective aperture
D 0.80 - 0.95
0.75
D 1.20 - 1.44 OIL
1.10
9.2.4 Oblique Illumination (TL)
• First adjust transmitted light darkfi eld.
• To obtain a relief-like contrast:
Condenser
UCA/P:
Rotate condenser disk slightly out of the
DF
po-
sition.
Condensers CL/PH, CLP/PH and
APL. ACHR.0.9 (P):
Push
DF
light ring slide part way in.
9.2.5 Polarization (TL)
• Swing the lambda plate of the lambda plate
compensator out if necessary.
• Insert a specimen and rotate an appropriate
objective into place.
• Bring the image into focus and set the Köhler
illumination.
• Insert the analyzer as far as the click stop into
the left side of the stand (Fig. 62). The engrav-
ing
λ
must be on the underside of the stand.
When using the Pol intermediate tube*:
Switch on the analyzer.
• Push the polarizer with the labeled side facing
upward
into the lower opening.
Fig. 61
Darkfi
eld
condensers
1
Upper part (dry)
2
Lower
part
3
Orientation
pin
4
Upper part (oil immersion)
1
2
3
4
Fig. 62
Polarization
1
Analyzer
2
Polarizer
1
2
Содержание DM2700 M
Страница 1: ...Leica DM2700M Instructions Bedienungsanleitung ...
Страница 3: ...Leica DM2700M Instructions ...
Страница 18: ...18 4 Overview of the Instrument 410 496 330 115 Dimensions specification in mm ...
Страница 79: ...Leica DM2700M Bedienungsanleitung ...
Страница 94: ...18 4 Geräteübersicht Abmessungen Angabe in mm 410 496 330 115 ...
Страница 155: ......