9
Data, Setting the mains voltage
Some of the following data may vary according
to the model:
Built-in supply Type L 30 T
(Fig. 1a)
Non-stabilized (major fluctuations in the mains
voltage may lead to visible falloff of image
brightness and colour shifts in photography).
Max. output
30
W (
%
12 V). Power consumption
45 VA.
Switchable voltage setting 230 V or 115 V (+10/
– 20 %) on the back of the microscope (1 and 2).
For users where the local mains voltage is
100 – 120 V:
the microscope has been set at
230 V at the factory! Make sure to disconnect
from the mains before switching the voltage
(230 or 115 V)!
Attention:
4
7
9 10 8
5
7 9 10
8
6 7 9 10 8
3
Attention:
Transit protection
To avoid damage caused by vibration, take the
following precautions before transporting the
microscope: disassemble the objectives and
condenser (14.7). Remove the stage after
loosening the clamp screw (28.13). Block the
vertical movement of the stage bracket by fitting
hard foam rubber between the microscope
base, stage bracket and objective nosepiece.
Built-in power supply
The DM L microscope series can be supplied
with 3 different built-in power supplies for 12 V
halogen lamps, depending on its intended
application.
Power supply
Fig. 1a – c
Backs of DM L microscopes
a
Type
L 30 T, b
Type
L 30 S, c
Type
L 100 S
1
Type plate,
2
115/230 V switch (type L 30 T only),
3
Connection for 12 V 100 W lamp, type L 100 S only; some configurations also
have 2 100 W ports with switch, e.g. for alternate operation (transmitted/incident light),
4
12 V 30 W halogen lamp,
5
Cover, only
for 30 W (removed in Fig. 3a),
6
Adapter ring with clamp screw for 100 W lamphousing,
7
Ground connection,
8
Mains switch
with pilot lamp,
9
Mains cable connection,
10
Fuse module with 2 fuses (type indicated on adhesive label)
a
b
c
1
2
1
1
!
Содержание DM L