Oblique illumination*
To obtain a relief-like contrast; push DF slide
(CL/PH and CLP/PH condenser; 7.7) in part way
or rotate condenser disc (18.15) slightly out of
the DF position (open aperture diaphragm).
Operation
of oblique illumination, polarization
Fig. 32
Accessories for contrast techniques*
1
Filter holder,
→
Fig. 5b and 33.4,
2
Transmitted light polarizer
IC/P,
→
(34.9 and 6.2),
3
Polarizer, Ø 32 mm,
→
(33.5a/b),
4
Compensator Ø 32 mm, in the example
λ
compensator, for
mounting in filter holder (33.6a) or in polarizer IC/P (34.4),
5
Light filter Ø 32 mm, for filter holder (33.1), and for incident
light universal illuminator,
6
Analyser,
→
(33.1),
7
Com-
pensator for tube slit
→
(33.8), in the example
λ
compensator,
8
IC prism, for tube slit (33.8, 34.6), in the example for
objectives with pupil position D,
→
p. 50 and 28,
9
Light ring
slide for CL/PH and CLP/PH condensers,
→
p. 42 – 44
Assembly of polarizers*
Push the
analyser
(32.6) into the microscope
stand (33.1). The engraving
λ
must be on the
underneath. When incident light illumination is
switched on, the analyser must be pushed into
the slot of the illuminator (on the left side). The
rotatable analyser* (not illustrated) is turned to
the zero position.
There is also an optional intermediate tube Pol*
(Pol module A/BL) with switchable analyser and
Bertrand lens (not illustrated).
Fig. 33
Polarization contrast*, transmitted light
1
Analyser,
→
(32.6),
2
Clamp screw for stage rotation,
3
Aperture diaphragm,
4
Filter holder,
→
Fig. 32.1,
5a, b
Polarizer, Ø 32 mm,
6a, b
λ
- or
λ
/4 compensator, Ø 32 mm,
7a, b,
c
Switch lever for transmitted light filter magazine,
→
Fig. 4,
8
Mount for IC prisms and compensators (with dust
protection flap)
1
2
3
5b
6b
4
5a
6
7a
7b
7c
8
b
1
2
3
4
5
6
7
8
9
a