Technical Data
Message Codes
Error message
appears beside the message number on the
display.
In case of this message switch on/off instrument
several times and check if message is still
displayed. Then call service and specify the
message displayed.
Care
Look after the optical surfaces with the same care that you
would apply to spectacles, cameras and field glasses.
*
Maximum deviation occurs under unfavorable condi-
tions, such as in bright sunlight or when measuring to
inadequately reflecting surfaces. For distances over
30m-without using a target plate- themaximum devia-
tion may increase by ±0.1mm/m to maximum
± 10mm. For more information, refer to
www.disto.com.
** As of 100m the measuring units are displayed in cm.
All illustrations, descriptions and technical specifica-
tions are subject to change without prior notice.
Measuring accuracy
(2x standard deviation)
DISTO™ special
5
typ.: ± 1,5mm / max.: ± 3mm*
Smallest unit displayed
1mm**
Range (from about 30m, use
target plate)
0.2m up to 200 m * **
Time for a measurement dist / trc
0.5...ca.4s / 0.16...ca.1s
∅
Laser spot (at distance)
6 / 30 / 60 mm (10 / 50 / 100 m)
Illumination
Multiline display
Multifunctional bumper
Time delay release
Pocket calculator
Tracking
Constant
10 values
min./ max. Tracking.
Pythagoras
Mermory (Stack)
15 Werte
Battery, Type AA, 2x 1.5V
up to 10’000 measurements
(only with Alkaline batteries!)
Splash and dust
proof
IP54 acc. IEC529: splash proof,
dust proof
Dimension
172 x 73 x 45 mm
Measuring accuracy of bubble
1°
Temperature range
Storage
Operating
-25°C to +70°C
(-13°F to +158°F)
-10°C to +50°C
(-14°F to +122°F)
Message
Code
Cause
Remedy
203
Wrong sequence with
Pythagoras meas.
Carry out measurement in
correct sequence
204
Calculation error
Repeat procedure
252
Temperature too high,
above 50°C (measuring)
Cool down instrument
253
Temperature too low,
below -10°C (measuring)
Warm up instrument
255
Receiver signal too weak,
measurement time too
long, distance <200 mm
Use target plate
measurement time >10
sec.
256
Received signal too
powerful
Use target plate (correct
side)
257
Wrong measurement,
ambient brightness too
high
Use target plate
260
Laser beam interrupted
Repeat measurement
All other messages
Call service "System"