16
Check & Adjust
The check and adjustment of the BLK3D is possible thanks to the
Check &
Adjust
function.
Check
- Is the process of verifying the accuracy of the device with the Initial
Accuracy Test.
☞
Can be done hand-held.
☞
Requires the BLK3D calibration plate.
☞
Is recommended after a fall.
Adjust
- Is the process of adjusting the accuracy of the device.
☞
Requires the BLK3D calibration plate, tripod and tripod adapter.
From the menu it is possible to:
•
See when the last check was done.
•
See when the last adjustment was done.
•
Check the accuracy.
•
Reset the adjustment status back to default settings.
To perform the Initial Accuracy Check
1.
Place the BLK3D calibration plate on the floor or against the
wall and point the laser at the centre of the calibration
plate.
2.
Fit the calibration plate into the red frame shown on screen.
Use the guidelines on the screen to adjust the distance of
the camera from the calibration plate (1.2 m/3.9 ft). Trigger
the picture using the
Photo
button on screen.
3.
After the check is complete, feedback is provided to the
user:
Check & adjust
50
Check & Adjust