background image

Cleaning, disinfection and sterilization of attachments:  

(except Rechargeable Battery 22/4415-00; 22/4416-00 and Desktop Charger 22/4413-00, 22/4414-00 or 22/4411-00, 22/4412-00 because these 
products are not categorized as 

 

medical devices ) 

 

 

CLEANING AND DISINFECTION:  

 

CARE AND MAINTENANCE  

 

Care and maintenance are vital to keep the attachments in good condition and fully functional! 

 

Causes of corrosion with stainless steel used for instruments. 

 

Corrosion is damage or where that can be caused by a chemical reaction, e.g. due to: 

 

 

1.) Damaged surfaces 

 

2.) The effect of surgical exudates: blood, suppuration or bodily secretions in prolonged contact with instruments. 

 

3.) Excessive exposure to certain solutions: saline and iodine solutions, chloride and strong acids, alkaline solutions and incorrectly 

 

used disinfectants 

 

4.) Poor water quality when cleaning, steam-sterilizing or rinsing instruments and attachments: e.g. due to rusty water pipes or 

 

penetration of rust and metal particles into steam sterilizers. 

 

5.) Inadequate maintenance of instruments and accessories: If rust forms, it may be transferred to other instruments and 

 

attachments (avoid contact at all costs because it is very dangerous for sterilization). 

 

 

Constant use of instruments and attachments constitutes natural wear, which shortens the service life correspondingly. 

 

Instruments and attachments that are used often should be replaced at regular intervals. 

 

 

MEASURES TO PREVENT CORROSION:  

 

1.) Careful laying out of correctly packaged and sterilized instruments before the operation: selection according to the type of i n - 

 

tervention. Always ensure that the outer packaging is intact and the expiry date has not elapsed. 

 

2.) Check the sterilization indicator inside the tray to ensure that the contents have been sterilized. 

 

3.) Arrange the instruments and attachments according to the sequence of use during the operation: if any instruments are not 

 

needed, leave them in the tray; only prepare instruments and attachments a short time before the operation. 

 

4.) Remove blood and other exudates from instruments and attachments during the operation: replace instruments and attach- 

 

ments. in their assigned place after use. 

 

5.) Flush through cannulated instruments to prevent blood and bone powder from drying on. 

 

6.) Use Ringer's lactate solution or saline solution to clean instruments and accessories, but do not immerse them in solutions. 

 

7.) Start cleaning immediately after the operation: instruments and attachments used during the operation are regarded as conta- 

 

minated. Instruments used on patients with possible infections must be similarly disposed of; wipe off any blood, bone powder and 

 

tissue remnants. 

 

8.) Clean instruments in a designated place within the surgery department. If the instruments are sent directly to the central ste- 

 

rilization area, first cover them up to prevent any risk of contamination for staff or the environment. Wear protective clothing when 

 

disposing of contaminated instruments and attachments! 

 

 

DISMANTLING FOR CLEANING:  

 

Instruments and attachments with removable parts should be dismantled. Keep all screws, nuts, bolts and other small parts together, 

 

ensuring it is clear which instrument they belong to. The persons responsible for dismantling and reassembling instruments should 

 

be appropriately trained. Clean the individual parts to remove any remaining blood or bone powder and prevent drying on during 

 

sterilization and consequent irreparable damage. 

 

 

DISINFECTION:  

 

When using chemical disinfectants, the manufacturer's instructions must be strictly followed: but always dilute solutions before use. 

 

For thermal disinfection with hot water, the water must be free of foreign bodies. 

 

Please note:

 Instruments and attachments made of anodized aluminium should not come into contact with certain disinfectant so- 

 

lutions or cleaning agents, so always check the solution before use to avoid subjecting the metal to a chemical reaction. 

 

 

CLEANING:  

 

Whether cleaning mechanically or manually, check carefully which cleaning agent is used with which 

 

method. Always observe the instructions for dilution and use.

 Recommended pH: between 7.0 and 8.5. 

10 

Содержание 22/4400-00

Страница 1: ...3 BATTERY DRILL SYSTEM User Manual Easy to use Ergonomic design Simple compact battery packs with high capacity Small robust drill with an extensive range of attachments Steam sterilisable Date 10 10 2011 ...

Страница 2: ...rse rotation hold down both triggers R and L Specifically reverse rotation is required for procedures such as releasing bone screws and removing Kirschner wires Reverse rotation must be used exclusively for the above mentioned procedures Note The Rechargeable Battery 22 4416 00 may not be sterilized but must be treated with the same care as a surgical instrument even though it is not categorized a...

Страница 3: ...ms Today health professionals and patients are at constantly increasing risk so appropriate cleaning and disinfection of these drills and attachments is extremely important If in doubt contact LEIBINGER GmbH Co KG and the manufacturer of the disinfectant and cleaning agent used Note Please observe the Instructions for use of attachments with the exception of the Rechargeable Battery 22 4415 00 22 ...

Страница 4: ...to manufacturer for re No sound when trigger is pressed R or L trigger pressed Remark The lower trigger has a double function main switch and re verse rotation upper trigger clockwise rotation pair Drill cannot be switched from Solenoid switch trigger defec Reversing function does not Send to manufacturer for re reverse to clockwise rotation tive Control system defective react Continuous running o...

Страница 5: ...f risk analysis Evaluation of the check list of Fundamental Requirements for this medical device Attachments Description Battery Drill System with end cap Mono Block Battery oscillating Saw Long rechargeable battery with extension Standard rechargeable battery Funnel for inserting rechargeable battery into Battery Drill System 4 807 Desktop Charger mono 240 115 V charging time 60 minutes Desktop C...

Страница 6: ...nce with the provisions of MED UMW LEIBINGER GmbH Co KG will deal with disposal Key to the symbols Symbol 1 Observe instructions in User s Manual Symbol 2 Classification Type BF Symbol 3 Recyclable Symbol 4 CE 1253 in accordance with Directive 93 42 EEC Symbol 5 Protection class II device Symbol description R clockwise rotation L reverse rotation 6 ...

Страница 7: ...use freshly charged batteries Preparation Step 1 Open the end cap 4 by turning in the direction indicated by the arrow Step 2 Insert the rechargeable battery 3 into the battery compartment 2 in the direction shown by the arrow at an angle of not more than 10 to 15 insert gently to avoid damaging the contacts Step 3 Replace the end cap 4 and close it by turning in the direction indicated by the arr...

Страница 8: ...y the arrow Slide the chuck 2 onto the shaft 3 as far as it will go Release the chuck collar 1 Turn the chuck slightly until it engages in the shaft 3 Check the connection Hold the chuck 4 with your left hand and pull it away from the drill Fault Clutch does not engage in shaft Rectification If due to dirt clean the clutch if due to damage send to Customer Service 1 3 4 2 8 ...

Страница 9: ... the Battery Drill System If it is not used there may be a loss of sterility Desktop Charger mono 240V and 115V charging time 60 minutes Please state required voltage when ordering Desktop Charger dual 240V and 115V charging time 60 minutes Please state required voltage when ordering Three jaw chuck 0 5 mm to 7 0 mm capacity for round and triangular shanks incl chuck key Reduction gears with AO st...

Страница 10: ...nd attachments a short time before the operation 4 Remove blood and other exudates from instruments and attachments during the operation replace instruments and attach ments in their assigned place after use 5 Flush through cannulated instruments to prevent blood and bone powder from drying on 6 Use Ringer s lactate solution or saline solution to clean instruments and accessories but do not immers...

Страница 11: ...packaging then drying is not necessary 5 LUBRICATION The joints and all moving parts of instruments and attachments must be cleaned and lubricated after use Only use biological oils and acid free lubricants pure cooking oil is recommended If special instrument oil solutions are used the instruments and attachments can be immersed for a certain period of time With some instrument washing machines t...

Страница 12: ...harged the lamp glows green 2 Chargers 22 4413 00 22 4414 00 or 22 4411 00 22 4412 00 switch off automatically when the BATTERY 22 4415 00 22 4416 00 is fully charged This means that you can leave the BATTERY 22 4415 00 22 4416 00 in the operational Charger 22 4413 00 22 4414 00 or 22 4411 00 22 4412 00 for a prolonged period so that it is always charged to full capacity without any risk of overch...

Страница 13: ...obtain a list detailing the type and scope of work performed with the company stamp date and signature of the repairer WARNING All equipment must be cleaned before sending for servicing Transport and storage conditions To prevent damage The Charger 22 4413 00 22 4414 00 or 22 4411 00 22 4412 00 and BATTERY 22 4415 00 22 4416 00 must be transported in the aluminium case provided which has a special...

Страница 14: ... EN 529 Chargers 22 4413 00 22 4414 00 or 22 4411 00 22 4412 00 240 V and 115 V 50 HZ 14 VA 12 V 1100 mA max II IPX0 Function The Chargers 22 4413 00 22 4414 00 or 22 4411 00 22 4412 00 are controlled and monitored by a microcontroller Irrespective of the charge level of the battery the controller measures the momentary capacity and controls the charging process 14 ...

Страница 15: ...Notice 15 ...

Отзывы: