LeCroy WAVESURFER 44MXS-B Скачать руководство пользователя страница 3

T.O. 33K3-4-3760-1

 

2  EQUIPMENT REQUIREMENTS: 

 
 

  

Minimum 

Use 

Calibration 

Sub- 

 Noun 

Specifications Equipment 

Item 

 
2.1  OSCILLOSCOPE  

Range:  0 to 

30 VDC 

Fluke 

 CALIBRATOR 

 

9500B/3200AF 

  

Accuracy: 

 

0.475% of setting 

 
 

 

Range:  Leveled Sinewave: 

 

 

600 to 720 mV p-p,  

 

 

10 kHz to 400 MHz 

 
 

 

Accuracy:  Flatness, 

5%  

 

 

relative to 50 kHz; 

  

Frequency, 

2.5 ppm 

 

3  PRELIMINARY OPERATIONS: 

 
3.1  Review and become familiar with the entire procedure before beginning the Calibration Process. 
 

WARNING 

 
Unless otherwise designated, and prior to beginning the Calibration Process, 
ensure that all test equipment voltage and/or current outputs are set to zero (0) or 
turned off, where applicable.  Ensure that all equipment switches are set to the 
proper position before making connections or applying power.  If not strictly 
observed, could result in injury to, or death of, personnel or long term health 
hazards. 

 
3.2  Connect the test equipment to the appropriate power source.  Set POWER switch to ON and allow a warm-up 
period as required by the manufacturer. 
 

3.3  Connect the TI to the appropriate power source.  Press the TI 

 switch and allow a 20 minute warm-up period.  

 
3.4  Throughout the Calibration Process, the TI front panel push-buttons and knobs appear in 

bold

, the touch screen 

menu selections in 

bold italicized

, and the icon and function softkeys in 

italicized

 font. 

 
3.5  Press TI 

Touch Screen

 to enable the touch screen as needed during the Calibration Process. 

 
3.6  When numeric entries are required, access a soft keypad by pressing the appropriate TI function softkey twice. 
 
3.7  Perform only those portions of the procedure that pertain to TI being calibrated. 
 
3.8  Unless otherwise specified, the TI controls should be set for several cycles of stable display with ample 
amplitude. 
 

NOTE 

 
 

The Active Head (p/o Oscilloscope Calibrator) is an integral part of the  

 

Oscilloscope Calibrator.  All connections are to be made through the Active  

 Head. 
 
 

Содержание WAVESURFER 44MXS-B

Страница 1: ...r contractors for official use or for administrative or operational purposes only 28 February 2014 Requests for this document shall be referred to AFMETCAL 813 Irving Wick Dr W Heath OH 43056 1199 Destruction Notice For unclassified limited documents destroy by any method that will prevent disclosure of the contents or reconstruction of the document Published under Authority of the Secretary of th...

Страница 2: ...upled 2 to 99 mV div 1 V 100 mV div to 1 V div 10 V 1 M coupled only 1 02 to 10 V div 100 V Accuracy 1 5 of offset 0 5 of FS 1 mV Bandwidth Range 2 mV to 1 V div 50 coupled Apply a constant amplitude DC to 400 MHz signal while changing frequency Vertical Accuracy Down not more than 3 dB deflection compared against deflection at a referenced frequency Trigger Range All Channels Internal and Compare...

Страница 3: ...lth hazards 3 2 Connect the test equipment to the appropriate power source Set POWER switch to ON and allow a warm up period as required by the manufacturer 3 3 Connect the TI to the appropriate power source Press the TI switch and allow a 20 minute warm up period 3 4 Throughout the Calibration Process the TI front panel push buttons and knobs appear in bold the touch screen menu selections in bol...

Страница 4: ...RACY CALIBRATION 4 1 1 Set the TI controls as follows File Recall Setup Recall Default Horizontal s ns 1 00 s C1 Volts div 2 00 mV Bandwidth 20MHz Averaging 5 then OK C2 Volts div 2 00 mV Bandwidth 20MHz Averaging 5 then OK Vertical 3 Volts div 2 00 mV Bandwidth 20MHz Averaging 5 then OK Vertical 4 Volts div 2 00 mV Bandwidth 20MHz Averaging 5 then OK Measure Measure Setup Show Table checked P1 Me...

Страница 5: ...scilloscope Calibrator for the first value listed in the Applied column of Table 2 into 1 M Set the OUTPUT to ON Table 2 Volts Div Applied VDC Limits VDC 2 00 m 6 m 4 630 to 7 370 m 6 m 7 370 to 4 630 m 5 00 m 15 m 13 150 to 16 850 m 15 m 16 850 to 13 150 m 10 0 m 30 m 27 75 to 32 25 m 30 m 32 25 to 27 75 m 20 0 m 60 m 56 50 to 63 50 m 60 m 63 50 to 56 50 m 50 0 m 150 m 142 75 to 157 25 m 150 m 15...

Страница 6: ...rify the TI P1 mean C1 indication is within the corresponding values listed in the Limits column of Table 2 4 1 6 Set the Oscilloscope Calibrator OUTPUT to OFF 4 1 7 Set the Oscilloscope Calibrator for the next value listed in the Applied column of Table 2 into 1 M Set the OUTPUT to ON 4 1 8 Repeat steps 4 1 4 through 4 1 6 4 1 9 Set TI Vertical V mV to the next value listed in the Volts Div colum...

Страница 7: ... the Oscilloscope Calibrator from the TI 2 connector and connect to the TI 3 connector 4 1 19 Press TI Vertical 2 then Vertical 3 to turn C2 off and C3 on 4 1 20 Repeat steps 4 1 3 through 4 1 14 for TI C3 using C3 controls and settings 4 1 21 Disconnect the Oscilloscope Calibrator from the TI 3 connector and connect to the TI 4 connector 4 1 22 Press TI Vertical 3 then Vertical 4 to turn C3 off a...

Страница 8: ...ct the Oscilloscope Calibrator to the TI 1 connector 4 2 3 Set the Oscilloscope Calibrator Amplitude for 750 mV DC into 1 M Set the OUTPUT to ON 4 2 4 Press TI Clear Sweeps and allow at least 100 sweeps as indicated on the TI display before proceeding 4 2 5 Verify the TI P1 mean C1 indication is within 764 25 to 735 75 mV 4 2 6 Touch TI C1 set the Offset to 750 mV then touch Close 4 2 7 Set the Os...

Страница 9: ... 2 13 for TI C2 using C2 controls and settings 4 2 17 Disconnect the Oscilloscope Calibrator from the TI 2 connector and connect to the TI 3 connector 4 2 18 Press TI Vertical 2 then Vertical 3 to turn C2 off and C3 on 4 2 19 Repeat steps 4 2 3 through 4 2 13 for TI C3 using C3 controls and settings 4 2 20 Disconnect the Oscilloscope Calibrator from the TI 3 connector and connect to the TI 4 conne...

Страница 10: ...v 100 mV Coupling 50ΩDC Averaging 5 then OK Measure Measure Setup Show Table checked P1 Peak to peak C1 P2 Peak to peak C2 P3 Peak to peak C3 P4 Peak to peak C4 Vertical 2 press twice to turn channel off Vertical 3 press twice to turn channel off Vertical 4 press twice to turn channel off Close 4 3 2 Connect the Oscilloscope Calibrator to the TI 1 connector ...

Страница 11: ...s TI Vertical 1 then Vertical 2 to turn C1 off and C2 on 4 3 14 Touch TI Timebase Horizontal Setup RealTime then set the Time Division to 20 0 µs 4 3 15 Touch TI Trigger Trigger Setup Source C2 then Close 4 3 16 Repeat steps 4 3 3 through 4 3 11 for TI C2 using C2 controls settings and readouts 4 3 17 Disconnect the Oscilloscope Calibrator from the TI 2 connector and connect to the TI 3 connector ...

Страница 12: ...oupling 50ΩDC Averaging 10 then OK C2 Volts div 100 mV Coupling 50ΩDC Averaging 10 then OK Vertical 3 Volts div 100 mV Coupling 50ΩDC Averaging 10 then OK Vertical 4 Volts div 100 mV Coupling 50ΩDC Averaging 10 then OK Cursors Cursors Setup Horizontal Time X1 0 0 ns X2 0 0 ns Vertical 2 press twice to turn channel off Vertical 3 press twice to turn channel off ...

Страница 13: ...se 4 4 10 Repeat steps 4 4 5 through 4 4 8 verifying the TI C1 and cursors indications are within 216 to 284 mV when performing step 4 4 6 4 4 11 Repeat steps 4 4 9 and 4 4 10 for a TI trigger level setting of 250 mV verifying the C1 and cursors indications are within 284 to 216 mV 4 4 12 Set the Oscilloscope Calibrator OUTPUT to OFF 4 4 13 Disconnect the Oscilloscope Calibrator from the TI 1 conn...

Страница 14: ...connector 4 4 24 Press TI Vertical 3 then Vertical 4 to turn C3 off and C4 on 4 4 25 Set the TI controls as follows Trigger Trigger Setup Slope Positive Source C4 Close 4 4 26 Set the Oscilloscope Calibrator OUTPUT to ON 4 4 27 Repeat steps 4 4 4 through 4 4 12 for TI C4 using C4 controls and settings 4 4 28 Disconnect the test setup 4 5 TIME BASE ACCURACY CALIBRATION 4 5 1 Set the TI controls as ...

Страница 15: ...nector 4 5 3 Set the Oscilloscope Calibrator for a 600 mV p p leveled sinewave at 10 MHz into 50 Set the OUTPUT to ON 4 5 4 Verify the TI P1 freq C1 indication is within 100 to 100 Hz 4 5 5 Set the Oscilloscope Calibrator OUTPUT to OFF 4 5 6 Set all POWER switches to OFF or STBY Disconnect and secure all equipment CALIBRATION PERFORMANCE TABLE Not Required ...

Отзывы: