T.O. 33K3-4-3760-1
14
Trigger
Normal
Measure Measure
Setup…
Show
Table
(checked)
P1
Frequency
C1
Close
4.5.2 Connect the Oscilloscope Calibrator to the TI 1 connector.
4.5.3 Set the Oscilloscope Calibrator for a 600 mV p-p leveled sinewave at 10 MHz into 50
. Set the OUTPUT to
ON.
4.5.4 Verify the TI P1:freq(C1) indication is within -100 to +100 Hz.
4.5.5 Set the Oscilloscope Calibrator OUTPUT to OFF.
4.5.6 Set all POWER switches to OFF or STBY. Disconnect and secure all equipment.
CALIBRATION PERFORMANCE TABLE
Not Required