Tip Selection and Exchange
The PP011 probe is supplied with a rigid tip. The spring tips PK1-5MM-109 (0.8 mm) or PK1-5MM-124 (0.38 mm)
combine sharp points with on-axis compliance and can be ordered directly from LeCroy as optional accessories. All
three tip styles provide reliable connections under a wide range of physical interconnect situations. While lacking
the on-axis compliance feature of the spring tips, the rigid tip has a larger diameter and is more robust when
exposed to physical stress at the tip. Select the proper tip for your application needs.
To change the tip or replace it when damaged, carefully grip the outer most portion of tip and pull straight out,
along the axis of the probe, using needle nose pliers. Do not attempt to grip the plastic insulator with pliers when
removing the tip, as this will squeeze the tip, which will make it difficult or impossible to remove. Do not grip the
outer gold plated tube which the tip slides into. With the tip removed, align the replacement tip with the hole and
begin the insertion with the pliers. The tip may be fully seated by placing the probe against a hard surface and
gently applying pressure.
To remove tip, grip the spring-loaded portion of the tip
Do not apply pliers to the plastic insulator.
beyond the outer sleeve, and pull straight out.
§ § §