0HDVXULQJ6NLOOV
&RXSOLQJ
Coupling refers to the type of signal coupling at the input to the
trigger circuit. The trigger coupling can be selected independently
from the following options:
À
DC: All the signal’s frequency components are coupled to the
trigger circuit for high-frequency bursts or where the use of AC
coupling would shift the effective trigger level.
À
AC: The signal is capacitively coupled, DC levels are rejected
and frequencies below 50 Hz attenuated (< 10 Hz for the
LC564
AND
LC584 S
ERIES
).
À
LF REJ: The signal is coupled via a capacitive high-pass filter
network, DC is rejected and signal frequencies below 50 kHz are
attenuated (< 100 MHz for the
LC564
AND
LC584 S
ERIES
). For
stable triggering on medium- to high-frequency signals.
À
HF REJ: Signals are DC-coupled to the trigger circuit and a low-
pass filter network attenuates frequencies above 50 kHz. For
triggering on low frequencies.
À
HF: To be used only when necessary for triggering on high-
frequency repetitive signals > 300 MHz, with a maximum trigger
frequency of > 500 MHz (> 1 GHz for the
LC564
AND
LC584
S
ERIES
). HF is automatically overridden and set to AC when
incompatible with other trigger characteristics (and SMART
triggers). Only one slope available.
67(3
Press the corresponding menu
to select “
60$57
”. The
screen will display as shown next page:
Содержание Digital Oscilloscopes
Страница 1: ...Hands On Guide to LeCroy Color Digital Oscilloscopes Revision C April 1998 ...
Страница 21: ... ...
Страница 36: ... HWWLQJ 6WDUWHG Now press the corresponding to select REDEFINE These menus will appear ...
Страница 45: ... 0HDVXULQJ 6NLOOV 67 3 Press again The screen will then display ...
Страница 49: ... 0HDVXULQJ 6NLOOV The screen will now display ...
Страница 52: ... 3XOVH KDUDFWHUL DWLRQ 67 3 67 3 Press Then press to select 3DUDPHWHUV The screen will display 67 3 Next press ...
Страница 74: ... QDORJ 3HUVLVWHQFH The screen will then display 67 3 Press ...
Страница 93: ... 0HDVXULQJ 6NLOOV ...
Страница 105: ... 0HDVXULQJ 6NLOOV 67 3 67 3 Press Press the corresponding menu to select XDO from the Grids menu ...
Страница 116: ... QDO VLV ZLWK LVWRJUDPV The screen will now display 67 3 Press the corresponding menu for PARAMETER SETUP ...
Страница 127: ... 0HDVXULQJ 6NLOOV ...
Страница 149: ... 0HDVXULQJ 6NLOOV Note With 6WRUH HV selected if the test is a Pass then Trace 2 will be stored to floppy ...
Страница 175: ... 0HDVXULQJ 6NLOOV ...
Страница 177: ... 0HDVXULQJ 6NLOOV 67 3 Press again to display 67 3 Press and then ...