lemaitrefx.com
Cleaning Procedure
6
Note: Cleaning is recommended after every 25 litres of fluid.
1) Using a #2 Philips screw driver remove the 4 screws.Two
by the jog holder and two on the rear, above the fan. (Fig. 1)
2) Lift the rear of the lid up then slide it forward.
3) Using a pair of Needle-nose pliers remove the 2 Hold-
Down Springs; this will release the insulation cover and
allow access to the 4-Port Rapid Clean Vaporizer. Grasp a
spring and lift until the spring disengages from its retaining
hole. (Fig. 2)
4) Remove the Insulation Cover. (Fig. 3)
5) Remove the 4 nuts from the top of the 4-Port R-C-V using
a 7/16” wrench. (Fig. 4)
6) Remove all 4 nuts, lock washer and flat washers as well
as the aluminum top plate and gasket material. (Fig. 5)
7) By removing the silicone tube from the inlet tube, the
chamber can now be held over a container for cleaning.
8) Use a flat screwdriver to clean the 4-Port R-C-V by scrap-
ing away loose deposits. Brush out all of the ash material
from each of the 4 ports. A light coating is okay; the ports do
not need to be spotless. (Fig. 6)
9) Use a 5/64” drill bit to clean the nozzle. Ensure both
nozzles are cleaned of any debris.
10) Ensure the mating block surfaces are clean to allow for
proper heat transfer to take place before replacing the
chamber
Fig. 1
Fig. 3
Fig. 2
Springs
Fig. 4
Fig. 6
Fig. 5
Inlet
•
Make sure the main power cord is disconnected from power before at-
tempting to remove the outer cover
•
Make sure unit has cooled sufficiently to perform the operation