-17-
Alarm Type
Display
Speaker
Flash LED, Vibrator and Top Red LED
Dose Alarm
3 beeps
every second
FLASH LED
1 flash every second
VIBRATOR
1 vibration (**) every second
** An extra vibration is added if the speaker is disabled.
TOP LED
3 red flashes every second
Dose Warning
(See note)
Rate Alarm
2 long beeps
every second
Rate Warning
(See note)
Note :
Acknowledge by pressing
+
or
-
button longer than 3s.
5.1 Dose and Rate Alarms or Warnings
The DMC 3000 has 2 configurable dose alarms from
1 µSv to 10 Sv
and 2 configurable rate alarms from
10 µSv/h (*) to 10 Sv/h
.
(*) Please contact Mirion Technologies for a rate set point
between 1 µSv/h and 10 µSv/h and see § 9.1.
Dose and rate warnings
are acknowledgeable by
pressing either push button for longer than
3 seconds.
The rate alarm and warning occur when the rate
exceeds the set point. The rate alarm and warning
disappear when the rate decreases to 80% of the set
point.
The pictogram on the display remains visible in
Pause
mode until the next entry.
A
larms