Offset Adjustment
After you have calibrated your meter, you can use the
OFFSET
OFFSET
OFFSET
OFFSET
OFFSET
feature to make fine addi-
tions or subtractions to the reading, without affecting the calibration itself. You must have
mode =
quan
quan
quan
quan
quan
(4 digits) or
quant
quant
quant
quant
quant
(6 digits).
For example if your weighing structure is altered after calibration and you want to subtract
the effect of 37kg of extra metalwork which was welded to the hopper, you can easily do this
by entering a value of -37 in the offset value.
Lockout Switch must be OFF
Press 3 seconds
Set1
Set2
Output
Alarms
Digit
OK
Max/Min
Reset
1
2
3
4
Set1
Set2
Output
Alarms
Digit
OK
Max/Min
Reset
Use DIGIT button to select each digit in
turn, UP or DOWN buttons to increase or
decrease each digit’s value. If you want
to set a negative value, use DIGIT to
select the left hand digit, and press the
down button to go below 0 to activate the
- sign. Press OK when done.
Set1
Set2
Output
Alarms
Digit
OK
Max/Min
Reset
Done!
You may want to adjust a SCALE FACTOR value also, without affecting calibration. See
the separate SCALE page for this feature.
Press to accept
OFF
Circuit board ON
Press repeatedly until you see
OF.St
OF.St
OF.St
OF.St
OF.St
(4 digits) or
OFFSEt
OFFSEt
OFFSEt
OFFSEt
OFFSEt
(6 digits), followed
by the existing offset value.
(Default is
000.0
000.0
000.0
000.0
000.0
fo 4 digits or
000.000
000.000
000.000
000.000
000.000
for 6 digits)