background image

STANDARD USER MANUAL n° #0020-3 

SMU#0020-03 

 

Page 6/25 

CHAPTER 3.

 

SPECIFICATIONS AND DESCRIPTION 

 

3.1.

 

S

PECIFICATIONS

 

 

Models 

HYDROXYCHROM-100

 

HYDROXYCHROM-160

 

H

2

 Outflow 

@ 1013/20°C 

HYDROXYCHROM-100 = 100 Nml/min  

HYDROXYCHROM-160 = 160 Nml/min 

H

2

 purity 

Maximum Hydrocarbon content: 0.1ppm 

Dew point 

-40°C / -40°F 

Outlet Pressure 

From 0.5 to 7 bar (7 to 102 psig), adjustable by software.  

Pressure resolution 

10 mbar 

Pressure stability 

Better than ±10 mbar 

H

2

 generation technology 

Proton Exchange Membrane (PEM), Solid Polymer Membrane 

Drying technology 

No maintenance static dryer 

Water quality 

High purity distilled & filtered water. TOC free. 
Conductivity < 0.20 µS/cm 

Water capacity 

5L tank outside, 0.4L tank inside.  

Water consumption 

5L water generates about 6000L Hydrogen 

Safety 

Low  H

2

  stored  volume;  over  pressure  valve;  internal  leak  test;  automatic 

shut down; maximum current limit, water quality. 

Manual control 

ON-OFF power switch 

Display 

By 

Hydroxychrom Viewer

 

Communications 

USB 

H

2

 outlet fitting 

Stainless steel 1/8" OD compression  

Functioning conditions : 

 

Temperature 

 

Humidity 

 

+10°C to +35°C 

max 80%, non condensing 

Transport 

and 

storage 

conditions: 

 

Temperature 

 

Humidity 

 

Duration 

 

 

+4°C to +40°C 

Max 90%  

Maximum 30 days. Instrument should run 5 minutes every month 

Power supply 

Automatic switching from 90VAC to 260VAC, 47 to 63 Hz 

Power consumption 

(max at full flow) 

HYDROXYCHROM-100 and HYDROXYCHROM-160: maximum 150W 

Sound pressure 

< 40dB (A) 

Содержание HYDROXYCHROM

Страница 1: ...e Application date Modified chapters 0 Creation 24 10 2012 All 1 Revision 25 10 2012 Ch 3 4 2 Revision 22 04 2015 Ch 4 6 2 3 Revision 14 12 2018 Ch 3 1 4 3 Editor S GUTIERREZ JP AMIET K TRAULLE Checke...

Страница 2: ...IONING PRINCIPLE 8 CHAPTER 4 INSTALLATION AND OPERATION 9 4 1 RECEIPT OF INSTRUMENT AND CHECK 9 4 2 DELIVERY CONTENT 9 4 3 GENERATOR INSTALLATION 9 4 4 CONNECTIONS 10 4 4 1 Fluidic connections 10 4 4...

Страница 3: ...the delivery date The replacement of the defective components will be free except the transport or travel fees that will be invoiced according the current tariffs CHROMATO SUD accepts no liability fo...

Страница 4: ...operations alarms as well as troubleshooting It is applicable to the below generation models and software HYDROXYCHROM Viewer Version 1 0 produced since October 2012 HYDROXYCHROM 100 HYDROXYCHROM 160...

Страница 5: ...cause 2 2 W E E E PRODUCT RECYCLING DECLARATION In agreement with the European EC 2002 96 directive on electrical and electronic equipment recycling this product may not be disposed in the garbage Fo...

Страница 6: ...d water TOC free Conductivity 0 20 S cm Water capacity 5L tank outside 0 4L tank inside Water consumption 5L water generates about 6000L Hydrogen Safety Low H2 stored volume over pressure valve intern...

Страница 7: ...ear panel left side has Mains plug with fuse and switch DB9 interface connectors for RS485 remote control USB dry relay The rear panel in the center has ZEROWATER input from external tank fitting for...

Страница 8: ...YDROXYCHROM 100 and HYDROXYCHROM 160 are equipped with a single layer H2 cell H2 GENERATION MODULE O2 Outlet Water Inlet H2 Outlet External 5L Water Tank ZERO WATER box Water pump 1 Static dryer 24 H2...

Страница 9: ...charged to the customer It is well recommended to keep the wood box for future shipments 4 2 DELIVERY CONTENT Quantity Description 1 Hydrogen Generator HYDROXYCHROM with its factory communication opt...

Страница 10: ...at a maximum distance of 1 2 m from the generator It can nevertheless be placed at the same level or above the instrument The maximum difference in height between the generator and the ZEROWATER tank...

Страница 11: ...it was not possible verify that the cable use has a sufficient section and has a ground wire 3X1 mm2 minimum Make sure laboratory differential circuit breaker can absorb an inrush current of at least...

Страница 12: ...provides User Interface to control Gas Generator Network through USB interface 4 6 1 SPECIFICATIONS Requirements PC or Laptop under WinXP SP2 or higher Win Vista Win 7 At least 5Mb of free space on th...

Страница 13: ...you must log the computer in administrator Remark If you want work with HydroxychromViewer in Russian start by installing the Cyrillic code page procedure below and then come back here only available...

Страница 14: ...STANDARD USER MANUAL n 0020 3 SMU 0020 03 Page 14 25 Step 3 Choose destination folder or use default Step 4 Complete installation...

Страница 15: ...chrom Viewer V1 0 when the installation is complete you have to do a commit cf procedure below If you just installed Hydroxychrom Viewer V2 0 when the installation is complete this window will appear...

Страница 16: ...STANDARD USER MANUAL n 0020 3 SMU 0020 03 Page 16 25 CYRILLIC CODE PAGE INSTALATION Step 1 Run install_cyrillic_codepage exe Step 2 Select your preferred language for this installation...

Страница 17: ...tallation When the installation is complete this window will appear clic on yes the computer will restart automatically saving the modification you just made Remark After the computer restarted you ca...

Страница 18: ...EDURE FOR COMPUTER UNDER EMBEDDED WINDOWS Open the program named command prompt in the toollbar In this program write commitcf as shown in the picture below and press enter The program asks you to res...

Страница 19: ...STANDARD USER MANUAL n 0020 3 SMU 0020 03 Page 19 25 CHANGING LANGUAGE Step 1 Open Application partition Step 2 Open Hydroxychrom folder Step 3 Open Hydroxychrom_v2 ini file with Notepad...

Страница 20: ...ep 4 Modify the second line of the code LanguageFile Hydroxychrom_UK lng for English language LanguageFile Hydroxychrom_FR lng for French language LanguageFile Hydroxychrom_RU lng for Russian language...

Страница 21: ...R MANUAL n 0020 3 SMU 0020 03 Page 21 25 And open it again USAGE Step 1 Connect USB Step 2 Start HYDROXYCHROM Viewer Step 3 Adjust the COM port number that is associated with USB port Step 4 Left Clic...

Страница 22: ...f following Off No gas generation On Gas generation is running but will be stopped after power off Continuous On Gas generation is running but will be automatically started every time after power on R...

Страница 23: ...ure does not match the Set Pressure Timeout Alarm is time in seconds given to the instrument after start to reach the Set Pressure IMPORTANT NOTE For security reasons the Instrument stops generation w...

Страница 24: ...ter should be exchanged once every 3 years In order to keep the instrument in its original conditions please use exclusively distilled and filtered water 5 2 LED AND BUZZER INDICATIONS Green Red Buzze...

Страница 25: ...rewed correctly Volume between H2 generator and consumer too big Pressure value could not be reached on time Increase value of timeout Verify capacity and consumption reduce consumption of GC 3 H2 Cel...

Отзывы: