STANDARD USER MANUAL n° #0020-3
SMU#0020-03
Page 10/25
4.4.
C
ONNECTIONS
4.4.1.
F
LUIDIC CONNECTIONS
ZEROWATER Inlet:
Disconnect the tube assembly which connects the inlet and outlet of the water circulation circuit.
This assembly is made for transport.
Place the ZEROWATER tank close to the H
2
-generator
Connect the tube to the device. Lubricate the end of tube with distilled water and introduce it in
the fitting, (labeled ZEROWATER INLET), push it firmly and screw the nut.
REMARK
The ZEROWATER tank has been designed to be placed on the floor at a maximum distance of 1.2 m from
the generator. It can, nevertheless, be placed at the same level or above the instrument. The maximum
difference in height between the generator and the ZEROWATER tank is 1m.
Hydrogen Outlet
: Hydrogen pressure is available at the OUTLET H
2
output on the back of the instrument.
This outlet is equipped with a stainless steel Swagelok 1/8” fitting.
Oxygen outlet:
remove the 8mm tube from the transport tube assembly and connect it to the O
2
Outlet.
The O
2
outlet must be kept at atmospheric pressure and without restriction. Some droplets of water
condensation are sometime present at the tube’s end. This is normal; water could be collected by a small
plastic glass.
Air inlet:
Air inlet must be dry air at 3 bar.