- 29 -
TEST
-
Load distribution:
: it displays the active channels status (weight, load percentage on each channel,
enabled channels and/or channels in error). This menu can also be accessed directly from
the weight display by pressing the TEST key.
-
Input Test:
: ensure that for each open input
is displayed,
is displayed when the input is closed.
-
Output Test:
: setting
ensure that the corresponding output opens. Setting
ensure that the
corresponding output closes.
-
Analog Output Option Test:
: It allows the analog signal to range between the minimum and the maximum values
starting from the minimum.
: current output test.
: voltage output test.
-
Millivolt Test:
: it displays the response signal of each load cell expressed in mV with three
decimals.
-
Millivolt stored at zero setting (only if
=
):
: it displays the response signal of each load cell, expressed in mV with three
decimals (see section
TARE WEIGHT ZERO SETTING
)
EVENTS LOG
The instrument can store up to 50 events; the oldest records are overwritten.
-
: it displays the last 50 events stored, starting from the most recent one:
: zero-setting from the calibration menu, press ENTER to display the value set to zero.
: theoretical calibration, press ENTER to display the full scale set.
: real calibration, press ENTER to display the sample weight used.
: tare setting via the keypad, press ENTER to display the set value.
: load distribution error, press ENTER to display the weight value that triggered the
alarm, press ENTER again to display the difference between the load percentage and the
stored value.
: weight alarm, press ENTER to display the alarm type.
: load distribution deletion.
: equalization.
: modification or deletion of the manual selection of active channels.
-
:
it deletes stored events; confirmation is requested (
),press ENTER to proceed
or press ESC to cancel.
+
+