Varioshake Shaking Apparatus

7 / 34

1

Use of the Distillation Apparatus

Intended Use

1.1

The information in these operating instructions must by all means be carefully read and observed. Only then a perfect
functioning of the Shaking Apparatus can be guaranteed. The units may only be installed and operated by persons who
have made themselves familiar with these operating instructions. The frequency of the shaking motion can be set and is
electronically regulated. Laboratory vessels that are to be used on the Shaking Apparatus must be fixed safe-to-operate.
The maximum usable shaking frequency is also determined by the kind and weight of the load. Provide sufficient working
space in the vicinity of the unit to put down accessories in use safely.

Caution:
due to increased risk of injury never reach into the unit as long as it is still in motion.

Improper Use

1.2

LAUDA Varioshake Shaking Apparatus, operated in a laboratory, are no Medical Devices. They fall neither under national
nor international Medical Device Directives and have to be used and applied accordingly. The Shaking Apparatus may not
be used in potentially explosive surroundings and may neither be set up nor operated in laboratory areas with aggressive or
corrosive ambient.
By all means make sure to prevent the shaking procedure creating an explosive atmosphere during operation of the Shak-
ing Apparatus.

2

Warranty

For all laboratory apparatus and their accessories from LAUDA-GFL, there is a warranty claim, as well for spare parts,
repairs and modifications, carried out by LAUDA-GFL. In order to identify defective units, we require both model and
serial number on the nameplate, on the back of the Shaking Apparatus and, if applicable, a copy of the invoice.

3

Before Initiation

The information in these operating instructions must by all means be carefully read and observed. Only then a perfect
functioning of the Shaking Apparatus can be guaranteed. A free of charge guarantee repair cannot be granted for defects
due to improper installation or handling.

Safety precautions are marked with the following symbols

Read and observe the

operating instructions

Warning of hand injuries

Warning of dangerous
electrical voltage

General warning

Before maintenance and repair
disconnect the unit all-pole from
the electrical mains (pull the plug
from the socket).

Содержание VS 15 B

Страница 1: ...Operating Instructions Varioshake Shaking Apparatus VS 8 O VS 8 B VS 8 OE VS 8 BE VS 15 O VS 15 B VS 15 T VS 15 R VS 30 O ...

Страница 2: ...2 34 Varioshake Shaking Apparatus 09 2020 ...

Страница 3: ...ferent motions orbital reciprocating orbital rocking and rocking Shaking frequency and remaining running time are LC displayed during operation It is possible to run the units with Digital Control Unit via PC RS 232 interface optional available LAUDA Varioshake Shaking Apparatus are extremely silent and universally applicable suitable for both gentle moving of liquids and vigorous mixing Before in...

Страница 4: ...4 34 Varioshake Shaking Apparatus 09 2020 ...

Страница 5: ...re 11 9 2 10 Disposal of Old Units 12 11 Technical Data 13 Varioshake Shaking Apparatus VS 8 B VS 8 BE 13 11 1 Varioshake Shaking Apparatus VS 8 O VS 8 OE 14 11 2 Varioshake Shaking Apparatus VS 15 O VS 15 B 15 11 3 Varioshake Shaking Apparatus VS 15 T VS 15 R 16 11 4 Varioshake Shaking Apparatus VS 30 O 17 11 5 12 Circuit diagram 18 Varioshake Shaking Apparatus VS 8 OE VS 8 BE 18 12 1 Varioshake ...

Страница 6: ...6 34 Varioshake Shaking Apparatus 09 2020 ...

Страница 7: ...ed accordingly The Shaking Apparatus may not be used in potentially explosive surroundings and may neither be set up nor operated in laboratory areas with aggressive or corrosive ambient By all means make sure to prevent the shaking procedure creating an explosive atmosphere during operation of the Shak ing Apparatus 2 Warranty For all laboratory apparatus and their accessories from LAUDA GFL ther...

Страница 8: ...ical Data of this manual The electri cal connection must ensure an all pole separation of the Shaking Apparatus from the mains at any time The left knob of the unit only models VS 8 OE and VS 8 BE must be off position O The voltage on the nameplate at the back of the unit must be identical to the mains voltage If they are identical con nect the unit to the mains For further information please also...

Страница 9: ... atmosphere during opera tion of the Shaking Apparatus To initiate press switch 1 Displays 2 and 3 show the memorized set points for running time and shaking frequency Operation and display elements on the control panel 7 1 j To switch the Shaking Apparatus on and off k Display shows the preset running time and the remaining running time during operation of the Shaking Apparatus Display shows the ...

Страница 10: ...1 If the Shaking Apparatus is not used for a longer period it should be separated from the mains Remote control via PC optional RS 232 interface 7 3 Caution Before initiation please check the Shaking Apparatus assembly to make sure that any hazard is excluded The Shaking Apparatus must be operated within field of vision If the Shaking Apparatus is in remote control mode take special caution in the...

Страница 11: ...rom the mains socket Exchanging the fuses 9 1 The two fuses of the Shaking Apparatus are situated in a drawer at the back of the unit In order to check or to exchange these fuses press the fixing clips on both sides of the drawer and pull The fuses may only be exchanged against fuses of the same type Information on the fuse type can be found on the nameplate next to the unit plug and in the spare ...

Страница 12: ...harge company date signature 10 Disposal of Old Units LAUDA GFL will take responsibility within the scope of the legal directives for an environmentally sound handling and dis posal of all used LAUDA GFL units as of the production year 1995 that are returned to us free of charge and will have it materially recycled Before the unit is returned a legally binding declaration must be provided from the...

Страница 13: ...tion reciprocating 20 mm 20 300 min 1 up to 60 min or continuous operation Electrical connection 115 V or 230 V 10 50 60 Hz 115 V or 230 V 10 50 60 Hz Mains fuse Internal fuse 10 A F1 F2 500 mA T 10 A F1 F2 500 mA T Power 65 W 65 W Protection type class I IP20 I IP20 Surrounding conditions Use only inside buildings also in incubation and tempering rooms not in explosion endangered surroundings Use...

Страница 14: ...tal 10 mm 20 500 min 1 up to 60 min or continuous operation Electrical connection 115 V or 230 V 10 50 60 Hz 115 V or 230 V 10 50 60 Hz Mains fuse Internal fuse 10 A F1 F2 500 mA T 10 A F1 F2 500 mA T Power 65 W 65 W Protection type class I IP20 I IP20 Surrounding conditions Use only inside buildings also in incubation and tempering rooms not in explosion endangered surroundings Use only inside bu...

Страница 15: ...cating 30 mm 20 300 min 1 1 min to 99 59 hrs or continuous operation Electrical connection 115 V or 230 V 10 50 60 Hz 115 V or 230 V 10 50 60 Hz Mains fuse Internal fuse 10 A F1 F2 500 mA T 10 A F1 F2 500 mA T Power 65 W 65 W Protection type class I IP20 I IP20 Surrounding conditions Use only inside buildings also in incubation and tempering rooms not in explosion endangered surroundings Use only ...

Страница 16: ...peration rocking 3 degrees to horizontal 2 50 min 1 1 min to 99 59 hrs or continuous operation Electrical connection 115 V or 230 V 10 50 60 Hz 115 V or 230 V 10 50 60 Hz Mains fuse Internal fuse 10 A F1 F2 500 mA T 10 A F1 F2 500 mA T Power 90 W 90 W Protection type class I IP20 I IP20 Surrounding conditions Use only inside buildings also in incubation and tempering rooms not in explosion endange...

Страница 17: ...king amplitude Shaking frequency Timer orbital 32 mm 20 250 min 1 with platform frame 20 200 min 1 1 min to 99 59 hrs or continuous operation Electrical connection 115 V or 230 V 10 50 60 Hz Mains fuse Internal fuse 10 A F1 F2 500 mA T Power 90 W Protection type class I IP20 Surrounding conditions Use only inside buildings also in incubation and tempering rooms not in explosion endangered surround...

Страница 18: ...haking Apparatus VS 8 OE VS 8 BE 12 1 A1 Frequency control A2 Forked light barrier C1 Condenser F1 Fuse 500 mA inert F2 Fuse 500 mA inert F3 Fuse 400 mA H1 Operation pilot lamp M1 Drive motor R1 Potentiometer S1 Timer Timed operation up to 60 minutes and continuous operation ...

Страница 19: ...pparatus VS 15 T VS 15 R VS 30 O 12 2 Varioshake Shaking Apparatus VS 8 O VS 8 B VS 15 O VS 15 B 12 3 A1 Microprocessor controlled frequency control A2 Forked light barrier B1 Interference filter C1 Condenser F1 Fuse 500 mA inert F2 Fuse 500 mA inert M1 Drive motor ...

Страница 20: ...supply ge gr yellow green PE Protective earth bl blue N sw black L1 Electrical fuses 13 1 Model Power Power consumption at mains voltage Mains fuse F4 F5 VS 8 O VS 8 B VS 8 OE VS 8 BE 0 065 kW 0 3 A at 230 V 10 A max 16 A Model Power Power consumption at mains voltage Mains fuse F4 F5 VS 15 O VS 15 B 0 065 kW 0 3 A at 230 V 10 A max 16 A Model Power Power consumption at mains voltage Mains fuse F4...

Страница 21: ...5 Mains fuse on site Varioshake Shaking Apparatus in 230 V with mains supply 230 V N PE 50 60 Hz connected through 3 pole shock proof plug system Varioshake Shaking Apparatus in 230 V with mains supply 400 V 3 N PE 50 60 Hz connected through 3 pole shock proof plug system Varioshake Shaking Apparatus in 230 V with mains supply 230 V 3 PE 50 60 Hz connected through 3 pole shock proof plug system ...

Страница 22: ... 8 0015522 Ball bearing seat 9 A000097 Forked light barrier 10 0030096 Mains connection cable 11 0012045 Drawer 12 0012915 Fine fuse 0 5 A inert 14 A000112 Driving belt for model VS 8 OE 0017330 Driving belt for model VS 8 BE 15 A000073 Motor please order together with belt pulley 0015575 Belt pulley for model VS 8 OE 0015565 Belt pulley for model VS 8 BE 16 0013788 Frequency control 17 0014320 St...

Страница 23: ...Seat for ball bearing 9 A000097 Forked light barrier 10 0030010 Mains connection cable 11 0012045 Drawer 12 0012915 Fine fuse 0 5 A T 13 0012044 Unit plug 14 0026615 Driving belt for VS 8 O 0017330 Driving belt for VS 8 B 15 A000073 Motor 0015565 Belt pulley for VS 8 O 0013737 Belt pulley for VS 8 B 17 0014320 Stand Please always state model and serial no of the water still when placing an order f...

Страница 24: ... bearing 8 0012798 Condenser 9 0013778 RS232 interface 10 0030116 Mains suppression filter 11 0030010 Mains connection cable 12 0012045 Drawer 0012915 Fine fuse 0 5 A T 13 0012044 Unit plug 15 0012141 Geared motor 16 0026606 10 Studded disc d 10mm 17 0012847 Forked light barrier 18 0026604 Toothed belt 19 0013311 Stand 20 0014310 Rubber metal buffer Please always state model and serial no of the w...

Страница 25: ...metal buffer 8 0012798 Condenser 9 0013778 RS232 interface optional 10 0030116 Mains suppression filter 11 0030010 Mains connection cable 12 0012045 Drawer 0012915 Fine fuse 0 5 A T 13 0012044 Unit plug 15 0012141 Geared motor 16 0012847 Forked light barrier 17 0015485 Ball bearing seat 18 0026264 Ball bearing 19 0015484 Eccentric 20 0014311 Stand Please always state model and serial no of the wat...

Страница 26: ...5522 Seat for ball bearing 6 A000097 Forked light barrier 7 0013778 RS232 interface optional 8 0030116 Mains suppression filter 9 0030010 Mains connection cable 10 0012045 Drawer 0012915 Fine fuse 0 5 A T 11 0012044 Unit plug 13 A000112 Driving belt 14 A000073 Motor 0015531 Beltpulley for motor 15 0014320 Stand Please always state model and serial no of the water still when placing an order for sp...

Страница 27: ...6 0025407 Securing ring 7 0014310 Rubber metal buffer 8 0013778 RS232 interface optional 9 0030116 Mains suppression filter 10 0030010 Mains connection cable 11 0012045 Drawer 0012915 Fine fuse 0 5 A T 12 0012044 Unit plug 14 0026297 Flat spring Set complete with holders 15 A000070 Geared motor Studded disc pre assembled pos 17 to be ordered with the motor 16 0012847 Forked light barrier 17 002660...

Страница 28: ... ml Erlenmeyer flasks Part No A000027 for 100 ml Erlenmeyer flasks Part No A000028 for 200 ml Erlenmeyer flasks Part No A000029 for 250 300 ml Erlenmeyer flasks Part No A000030 for 500 ml Erlenmeyer flasks Part No A000031 for 1000 ml Erlenmeyer flasks Part No A000053 for 2000 ml Erlenmeyer flasks Clamps for separating funnels made of stainless steel Supplied com plete with fixing material to be sc...

Страница 29: ... gummierten Querstangen Part No A000050 with 6 rubber coated bars for VS 30 Adhesive mat black Strong adhesive special mat for easy attachment of different vessels onto the shaking platform or tray Max shaking speed up to 250 min 1 depending on shape and weight of the vessel as well as on the shaking amplitude Dimensions 200 x 200 mm the mat can be cut or trimmed with scissors Temperature 15 bis 5...

Страница 30: ...t shaking of the tubes Supplied complete with fixing material to be screwed onto a shaking tray Part No A000059 for 24 tubes 75 160 mm length Æ 12 17 mm Part No A000060 for 16 tubes 75 160 mm length Æ 25 29 mm Platform Frames to increase the capacity of the Shaking Apparatus made of stainless steel with 4 levels Part No A000051 for VS 15 T and VS 15 R ...

Страница 31: ...09 2020 Varioshake Shaking Apparatus 31 34 ...

Страница 32: ...32 34 Varioshake Shaking Apparatus 09 2020 ...

Страница 33: ...09 2020 Varioshake Shaking Apparatus 33 34 ...

Страница 34: ...LAUDA GFL Gesellschaft für Labortechnik mbH Schulze Delitzsch Straße 4 5 30938 Burgwedel Deutschland Tel 49 0 5139 9958 0 Fax 49 0 5139 9958 21 E Mail info lauda gfl de Internet www GFL de ...

Отзывы: