Varioshake Shaking Incubator
09/2020
Operation panel temperature controller
8
Display for set and actual temperatures
9
Pilot lamp "actual temperature" in the display
10
Pilot lamp difference value of over-temperature alarm in the display
11
Pilot lamp set temperature in the display
12
Change-over key between display and entry of higher set points
13
Entry of lower set points
22
Key to switch off alarm signal
23
Pilot lamp difference value of under-temperature alarm in the display
24
Change-over key for alteration of set points
25
Confirmation of new set points 9 Kontrolllampe Ist-Temperatur im Display
Operation panel shaking frequency control
14
Display for actual and set shaking frequencies
15
Pilot lamp "actual shaking frequency" in the display
16
Pilot lamp "set shaking frequency" in the display
17
Change-over key between the displays and entry of higher set points
18
Entry of lower set points
26
Shaking movement on/off
27
Change-over key for alteration of set points
28
Confirmation of new set points
6.3.1
Incubation time
Press key 20 of the keypad to switch the timer to entry mode.
The display begins to flash and pilot lamp 4 glows (display in minutes). The required incubation time can now by set by
pressing keys 6 (+) and 7 (-) (max. 59 minutes). Press key 21 to confirm the set minutes. The entry mode then switches
to entry of set hours. Pilot lamp 3 glows. Use keys 6 (+) and 7 (-) to set the required incubation time in hours (max. 999
hours). Press key 21 to confirm the set hours. The timer is now activated, the set incubation time starts to run down.
After activation of the timer the set incubation time can only be interrupted or altered by switching the Shaking Incubator
off with the main switch.
The currently remaining run-time is shown in the display. If the remaining run-time is more than one hour, pilot lamp 3
glows. The time is displayed in hours. If the remaining run-time is less than one hour, pilot lamp 4 glows. The time is dis-
played in minutes. While the incubation time is running down, press key 6 to choose between display of the remaining run-
time and the set total run-time. When display of the total run-time is chosen, pilot lamp 5 glows. Approx. 30 seconds
after the last entry, the digital timer switches back to normal operation mode. Expiry of the set incubation time is signalled
by a continuous acoustic alarm. Press key 19 to switch off the alarm signal. The timer merely triggers the alarm signal; the
unit is not switched off automatically.
Содержание Varioshake VS 60 OI
Страница 1: ...Operating Instructions Varioshake Shaking Incubator VS 60 OI ...
Страница 2: ...2 26 Varioshake Shaking Incubator 09 2020 ...
Страница 4: ...4 26 Varioshake Shaking Incubator 09 2020 ...
Страница 6: ...6 26 Varioshake Shaking Incubator 09 2020 ...
Страница 18: ...18 26 Varioshake Shaking Incubator 09 2020 Varioshake Shaking Incubator VS 60 OI ...
Страница 24: ...24 26 Varioshake Shaking Incubator 09 2020 ...
Страница 25: ...09 2020 Varioshake Shaking Incubator 25 26 ...