Alpha
9 / 52
(
(
Reset to factory default:
Display:
dEF
“Enter” pressed for longer than
3 seconds sets back to default:
SEt
= 25 °C
;
Hi
= 105 °C
;
Lo
= -30 °C
;
Cool
= Auto.
(
(
(
Calibrate probe (offset):
Input temperature of reference
thermometer (only offset values
±5 K of actual bath temperature are
accepted).
Display flashes. Set value
with or . Value is accepted
after 4 seconds or proceed immedi-
ately with the key: .
(
Flashing display “00.00” (preset)
for inactive timer or “hh.mm” for
time to initiate status “Standby”.
Activate timer by setting the Auto
Shut-Down Delay “hh.mm” with
or .
Deactivate timer by setting
“00.00”.
The setting is accepted only by
pressing the key: .
Bath temperature
2 % 3
Display flashes. Set Setpoint
with or . Value is accepted
after 4 seconds or proceed immedi-
ately with the key: .
Maximum Temperature
H i
Minimum Temperature
L o
Configure Refrig. System
C o o l
l
Offset probe
C A L
Settings default
d E F
Timer “Auto Shut Down”
A S d
End of menu
E n d
Setpoint
S E t
Alarm
n n n
Stand By
S t b Y
Standard
Alarm
Standby (timer event)
( blinking in timer mode)
Configure Refrigeration System
Default:
Auto
. Possible settings:
on
refrigeration on
Auto
refrigeration on if needed
oFF
refrigeration off
(3 sec.)
Содержание RA 12
Страница 12: ...12 52 Alpha 7 2022 1 Control head with clamp 2 Bath vessel 3 Four mounting feet 1 2 3...
Страница 14: ...14 52 Alpha 7 2022 1 Mains cable 2 Rating label 3 Cooler cable 4 Bath drain nozzle 1 2 3 4...
Страница 44: ...44 52 Alpha 7 2022 Cooling characteristics Alpha RA 8 RA 12 and RA 24 230 V 230 V...
Страница 45: ...7 2022 Alpha 45 52 11 Declaration of Conformity and Product Returns...
Страница 46: ...46 52 Alpha 7 2022...
Страница 47: ...7 2022 Alpha 47 52...
Страница 51: ......