Proline Kryomats
YAUE0007 / 2/26/2009
Starting up
53
7.6.3 Setpoint offset operating mode
With this function it is possible to apply an offset value to the temperature provided by the external
temperature probe or a module and then to use it as the setpoint. The bath temperature can, for
example, be operated at -25 °C below the temperature of a reactor, which is being measured by the
external temperature probe.
Command
−
Offset source
and
Setpoint offset
−
Open the device parameter menu via the
soft key Menu .
−
With the cursor keys continue to
Æ
Control
Æ
Setpoint offset
Æ
Offset
source
.
−
Off
indicates that the setpoint offset is
currently deactivated.
−
Select the setpoint source with
or
and confirm with
.
−
Interfaces (e.g. RS232) are only
displayed if a valid setpoint has already
been transmitted.
−
With the cursor keys continue to
Æ
Setpoint offset
.
−
The standard value is
0.00°C
.
−
Open the left-hand window with
.
−
Enter the desired temperature.
−
Accept the change with
,
−
quit the window with without making
changes.
Help Menu
End T
set
T
fix
Off
extern Pt100
RS232
Offset source
Setpoint offset
Help Menu
Screen T
set
T
fix
6 7 8 9
0
1 2 3 4
5
Input Setpoint offset:
0.
00
Min: -500.00°C Max: 500.00°C
Help Menu
End T
set
T
fix
0.00 °C
Offset source
Setpoint offset