Master: Alarm message
a
l
/B
Command: Pump blocked
(Pump motor monitoring: Overload,
blockage).
Ì
1. The viscosity of the heat teransfer liquid is too
high
Æ
change heat carrier liquid or raise the
setpoint temperature.
2. The pump is blocked.
Æ
Contact the LAUDA
Service Constant Temperature Equipment
Ì
Master: Alarm message
Os
l
DT
Command: Low level (pump)
(Pump motor monitoring: No load).
Ì
1. No liquid in the system. If this occurs, the level
monitoring has failed.
Æ
Check whether the float
in the expansion vessel is blocked by foreign
bodies. Otherwise, contact LAUDA Service
Constant Temperature Equipment
Ì
2. With the option "open load" the device draws air
out of the open load.
Æ
Move the return to the
load.
Master: Alarm message
Dppmp 00
Command: Overpressure
(outflow pressure too high).
Pump level too high
Æ
Select a lower pump level
Ì
Compressor overtemperature
Ì