5.4
Switching on the device
WARNING!
Overheating of the pump
Device damage
Never operate device without heat transfer liquid.
Personnel:
Operating personnel
1.
Switch on the device using the mains power switch.
A signal tone sounds. The actual temperature is shown
on the display. The pump is started. The refrigeration unit
is started after approx. 2 minutes. The blue LED lights if
the refrigeration unit is active.
2.
Depending on the size of the consumer, heat transfer liquid
must be refilled if necessary. Monitor the level indicator for
this.
Further information about refilling heat transfer
liquid can be found in
heat transfer liquid’ on page 35
5.5
Refilling heat transfer liquid
Personnel:
Operating personnel
Protective equipment:
Protective goggles
Protective clothing
Protective gloves
DANGER!
Use of incorrect heat transfer liquid
Fire
Select a heat transfer liquid with a temperature
range 20 K above the temperature range of the
application.
Commissioning
Microcool
35