Command
Description
IN_SP_05
Query of outflow temperature limit Lo
IN_PAR_00
Query of the control parameter Xp
IN_PAR_01
Query of the control parameter Tn (181 = OFF)
IN_PAR_02
Query of the control parameter Tv
IN_PAR_03
Query of the control parameter Td
IN_MODE_00
Keyboard master: 0 = free / 1 = blocked
IN_MODE_02
Standby: 0 = device ON / 1 = device OFF
TYPE
Query of the device type (answer = "LOOP")
VERSION
Query of the software version
STATUS
Query device status 0 = OK, -1 = fault
STAT
Query of fault diagnosis response: XXXXXXX; X = 0 no fault, X = 1
fault
1st Character = error
2nd Character = not assigned, always 0
3rd Character = not assigned, always 0
4th Character = not assigned, always 0
5th Character = sublevel
6th Character = not assigned, always 0
7th Character = not assigned, always 0
Take the following information into account:
" " (blank space) can also be used for "_".
If not otherwise specified in the command, the response will
always be given in fixed point format "XXX.XX" or for negative
values "-XXX.XX" or "ERR_X".
5.11.5
Error messages
The following is a description of the error messages of the RS 232
interface.
Error
Description
ERR_2
Wrong entry (for example, buffer overflow)
ERR_3
Wrong command
ERR_5
Syntax error in value
Operation
LOOP
34