There are two ways of checking:
n
Enter the ping command together with the IP address.
ping XXX.XXX.XXX.XXX
@
For "XXX.XXX.XXX.XXX", put the IP address that was entered
when the Ethernet interface was configured.
Or
n
Enter the ping command together with the serial number of the
thermostatic circulator (possible with software control system
1.36 or later).
ping serial number
@
If the Ethernet interface is configured and connected correctly,
the interface will return four responses within a very short time.
Fig. 21: Example for entering the ping command
It is easy to check the connection to the interface with a PC and Microsoft
Windows operating system.
n
For Windows 3.11 with the program "Terminal".
n
For Windows 95/98/NT/XP the program "HyperTerminal".
n
For operating systems Windows Vista, Windows 7, Windows 8 and
Windows 10 "HyperTerminal"* is not part of the operating system.
* Terminal programs are available on the Internet as freeware. These
programs offer features similar to "HyperTerminal" (for example
PuTTY or RealTerm). Search query "serial port terminal program".
Check the LAN network and the process
interface
V6
Integral Process Thermostats and High-Temperature Thermostats
49 / 198
Содержание IN 1030 T
Страница 89: ...6 3 Menu structure Fig 45 Menu Part 1 V6 Integral Process Thermostats and High Temperature Thermostats 89 198 ...
Страница 183: ...Fig 99 Declaration of Conformity V6 Integral Process Thermostats and High Temperature Thermostats 183 198 ...
Страница 184: ...Fig 100 Declaration of Conformity V6 Integral Process Thermostats and High Temperature Thermostats 184 198 ...
Страница 186: ...V6 Integral Process Thermostats and High Temperature Thermostats 186 198 ...
Страница 187: ...V6 Integral Process Thermostats and High Temperature Thermostats 187 198 ...
Страница 199: ......