Starting up
34
6.4.5.1 Setting the proportional band of the PID-controller
and 4
x
and
3x
−
Directly from level 0 (setpoint selection), press the key
combination on the left (top one on RE 3xx, bottom one
RE 2xx), until the PARA function is reached, then
−
switch to level 2 from PARA (see above) with key on the
left.
or
−
Within this level it is possible to scroll with the keys.
−
The display shows the current bath temperature Xp
and the current setting. To alter the setting press the
key on the left. Available settings from 0,5 to 9,9 °C .
(
➾
Section 6.4.5.2.
☞
Note).
(
➾
Section 9. Technical data))
−
Display flashes (approx. ca. 4 s).
or
−
During this time start to set the required value with one
of the two keys
−
Forward with the key to "Setting the reset time" or
−
with the key back to "PARA".