Starting up
36
RE 2xx), until the PARA function is reached, then
and
2x
−
switch to level 2 and move forward with key on the left
−
The display shows the current bath temperature and the
currently set interface. To alter the setting press the key.
or
−
Display flashes (approx. 4 s.). During this time start to
set the required interface.
−
Having chosen the RS 485 interface press the key.
−
The display shows the current bath temperature, ADR
and the current unit address (e.g.. 15).
Press the key, display flashes (approx. 4 s.)
or
−
During this time start to set the required unit address.
(possible setting: 000...127).
−
Forward with key to ”Setting the baud rate” or
−
back with key to ”Selection of the interface” resp. when
pressing the key twice you return to ”Setting the reset
time...”“
6.4.5.4 Setting the Baud rate/ transfer rate (serial interface)
and
4x
a
nd
3x
−
Directly from level 0 (setpoint selection), press the key
combination on the left (top one on RE 3xx, bottom one
RE 2xx), until the PARA function is reached, then