Technical data
26
9
Technical data (to DIN 58966)
Common technical data
E 100
Ambient temperature range
°C
5 to 40
Setting resolution
°C
0.1
Indication resolution
°C
0.1
Indication accuracy
°C
±0.4°C ± 0.5 % additive re-calibration from indicated value
➂
Temperature control
± °C
0,05
Safety features
➀
FL
Additional functions
Power
consumption
230 V;50/60 Hz
115 V; 60 Hz
100 V:50/60 Hz
kW
1,6
1,4
1,1
➀
FL: suitable for flammable and non-flammable liquids; NFL: only suitable for non-flammable liquids
➁
➾
Section 7.2. last item
Immersion thermostats
E 100
Operating temperature range
➀
°C
25 to 120
" with water cooling
°C
20 to 120
Operating temperature range
➁
°C
-20 to 120
Heater
power
230 V; 50/60 Hz
115 V; 60 Hz
100 V; 50/60 Hz
1.5
1,3
1,0
Pump type
pressure pump with choice of 5 output steps
Max. discharge pressure
➂
bar
0.4
Max. flow rate
➂
l/min
20
Pump connections
mm
nipples 13 mm dia.
Bath depth
➃
mm
min 150
Usable depth
➃
mm
min 100
Overall size (WxD)
mm
125x133
Height (H)
mm
315
Weight
kg
3
Ref. No.
230 V; 50/60 Hz
115 V; 60 Hz
100 V; 50/ 60Hz
LCE 0221
LCE 4221
LCE 6221
➀
at pump output step 1
➁
with additional cooling
➂
at pump output step 5
➃
baths
➾
section 10. Accessories
Units to EU-Directive 89/ 336/ EWG (EMC) and 73/ 23/ EWG (low-voltage) with CE-mark.
We reserve the right to make technical alterations!
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com