Through flow coolers
18
16/03/2018 / YAFE0004
7.5
DLK 45 LiBus and Proline heating thermostats with proportional cooling
method
The DLK 45 LiBus is switched on and off by the Proline heating thermostat.
Operation of the DLK control unit is only possible with Proline heating thermostats (no cooling
thermostats!).
The
“No.” specified on the rating plate located on the back of the master unit, must be ≥ 04-0001.
The use of a through flow cooler has to be defined in the thermostat.
The system operates with proportional cooling and automatic compressor.
The DLK only cools in case that the thermostat is switched ON.
Switches located on the DLK
DLK 45 LiBus
LiBus-cable
Mains ON
This method of operation has the same functions as the proportional cooling of LAUDA cooling
thermostats.
Proline thermostat with software version:
for control system V1.33
for operating system (command) V1.36
or higher required.
The thermostat allows switching between the following DLK modes:
"0" (always OFF),
"1" (always ON) and
"A" (automatic mode).
Master:
MNEnu
MNodu
Cool
Co
0
1
A
Command:
Menu
Interfaces
SmartCool
Cooling mode
off
on
automatic
DLK mains
cable