Alpha
June 20, 2018/ YACE0092
Starting up
25
7
Starting up
7.1
Connection to the supply
Compare the supply voltage with the data on the rating label (rearside on control head).
Connect the unit only to a grounded mains power socket (PE).
No warranty when the thermostat is connected to a wrong supply!
Please make sure that your mains plug is equipped with at least the following safety
fuses (
10).
Power supply Fuse protection
230 V~
12 A
115 V~
15 A
100 V~
15 A
The start current of the refrigerating machine may exceed those currents distinctly for
a short time.
Without external circuit ensure that the pump pressure outflow is linked to the pump
return.
Ensure that the unit is filled in accordance with Section (
6.2).
Note for electric installation on site:
The devices must be protected with a 16 ampere circuit breaker fitted during installation.
Exception: Devices with 13 ampere UK plugs.
7.2
Switching on
1 s
Switch on at the mains switch; an acoustic signal is
emitted for about one second.
Initial check of display and indicators; software version is
shown and the self-test becomes initiated.
The current bath temperature is shown on the display.
If necessary add more heat transfer liquid to
replace the amount pumped out to the exter-
nal circuit.
Bath temperature
2 % 3
Self-test
8 8 8 8
Содержание Alpha A
Страница 1: ...2SHUDWLQJ LQVWUXFWLRQV LAUDA Alpha HDWLQJ DQG FRROLQJ WKHUPRVWDWV...
Страница 2: ......
Страница 4: ......
Страница 15: ...Alpha June 20 2018 YACE0092 Control and functional elements 15 A 6 A 12 A 24 RA 8 RA 12 RA 24...
Страница 48: ...Alpha 48 June 20 2018 YACE0092 intentionally left blank...
Страница 50: ......
Страница 51: ......