9-Out of Range Notice 

When the head is exposed to a source exceeding its 
full scale, the LCD shows four short lines instead of a 
numeric value. 

 

10-Wavelength Selection

 

To select the wavelength of operation (e.g. from CO2 to 
YAG), go back to Paragraph 2 (shut the instrument off), 
but keep the 

ON/Zero

 button pushed until the wavelength 

change appears on the display (this step takes 

approximately 7 seconds). At this point release the button and the instrument is 
ready for a new measurement.. The displayed value is the last recorded power 
measurement carried out at this wavelength setting. 

 

11-Customer’s calibration 

This instrument can be re-calibrated by the user to 
match a custom reference. Remove the plastic cover of  
the adjustment switch located at the right side of the 
instrument body. Zero the instrument as indicated in 
Paragraph 2 of this Operation Section and afterwards 

carry out a measurement of a known laser source. Shift the switch on + or – if 
you need to increase or decrease the reading displayed on the LCD to match your 
reference. This is done by limited pushes on the 

ON/Zero

 key, until the desired 

power value is achieved on the display. Reset the switch to its middle position 

CAL

 and re-cap. The instrument is ready for new measurements.  

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 

 

Calibration 

 

 
Model: Fit ______                                         Serial number:  ______________    
 
Date of Calibration: _______________ 
 
Wavelength 1:      _______________nm 

 

Reference Power :________________W     Measured Power: _________________W 
 
Wavelength 2:      _______________nm 

 

Reference Power :________________W     Measured Power: _________________W 
 

 
 
 

 
LaserPoint certifies that, at the time of calibration, the above listed instrument meets or 
exceeds all technical specifications defined at page 8 of this Instruction Manual. 
 
 
The above listed instrument has been calibrated using standard references traceable to 
the National Institute of Standards and Technology (NIST) or Physikalisch Technische 
Bundesanstalt (PTB). 
 
 

This is not a Calibration Certificate for traceability purpose; such a Calibration 
Certificate  can be quoted on demand. 

 
 
 

 
Important - For Customer Use Only 
 

The reference setting of this instrument can be matched by user to a custom own 
reference by means of the instructions reported at Paragraph n.11 of the Operation 
Section of this Manual. 
 

 

 

 

 

 

Содержание Fit

Страница 1: ...Nov 2008 Firm vers 11 1 WARNING The user of this instrument must be trained to the use of power lasers and associated risks LaserPoint is in no way liable for any damage resulting from improper careless or above rated limits use of this instrument ...

Страница 2: ...dment 2 1997 Limits and methods of measurement of radio interference characteristics of information technology equipment Class B EN 61000 4 2 1995 Electromagnetic compatibility EMC Part 4 Testing and measurements techniques Section 4 2 Electrostatic discharge Class B EN 61000 4 4 1995 Electromagnetic compatibility EMC Part 4 Testing and measurements techniques Section 4 Electrical fast transient b...

Страница 3: ...technical specifications defined at page 8 of this Instruction Manual The above listed instrument has been calibrated using standard references traceable to the National Institute of Standards and Technology NIST or Physikalisch Technische Bundesanstalt PTB This is not a Calibration Certificate for traceability purpose such a Calibration Certificate can be quoted on demand Important For Customer U...

Страница 4: ...ely 2 seconds The instrument will also display one of the 2 available laser wavelengths e g CO2 The wavelength displayed the first time is random to change it refer to Paragraph 10 of this section 3 Auto check Cycle When the ON Zero button is released the instrument starts an auto check cycle during which both circuit and display are controlled and reset During this cycle the LCD shows all availab...

Страница 5: ... turns steady green The LCD shows the last measured power value Return to Par 5 for a new measurement 7 Cooling the Head Should the probe reach its limit temperature the LCD shows COOL and the LED blinks red Use spontaneous or air forced cooling Never use liquids As soon as the temperature returns below its maximum allowable limit the LCD displays the last measured power value Warning sudden cool ...

Страница 6: ... power measurement carried out at this wavelength setting 11 Customer s calibration This instrument can be re calibrated by the user to match a custom reference Remove the plastic cover of the adjustment switch located at the right side of the instrument body Zero the instrument as indicated in Paragraph 2 of this Operation Section and afterwards carry out a measurement of a known laser source Shi...

Страница 7: ...lf It is emphasised that this no cost return option is only available when you purchase a new product of equivalent type that fulfils the same function Please inform your supplier of your wish to have the old instrument collected when ordering the replacement If you wish to dispose of the instrument without replacing it then the appliance must not be mixed with unsorted municipal waste The crossed...

Страница 8: ...ack 8 Parameter Fit 50 Fit 200 Fit 500 Maximum Measurable Power W 50 200 500 Minimum Measurable Power 3 Accuracy W 1 5 1 5 1 15 1 Absolute Minimum Measurable Power W 1 25 3 7 5 Maximum Laser Beam Diameter mm 20 20 25 Power density damage threshold full scale at YAG wavelength W cm2 10000 Power density damage threshold full scale at CO2 wavelength W cm2 2500 Repeatability 1 ADC Resolution W 0 06 0 ...

Отзывы: