9-Out of Range Notice
When the head is exposed to a source exceeding its
full scale, the LCD shows four short lines instead of a
numeric value.
10-Wavelength Selection
To select the wavelength of operation (e.g. from CO2 to
YAG), go back to Paragraph 2 (shut the instrument off),
but keep the
ON/Zero
button pushed until the wavelength
change appears on the display (this step takes
approximately 7 seconds). At this point release the button and the instrument is
ready for a new measurement.. The displayed value is the last recorded power
measurement carried out at this wavelength setting.
11-Customer’s calibration
This instrument can be re-calibrated by the user to
match a custom reference. Remove the plastic cover of
the adjustment switch located at the right side of the
instrument body. Zero the instrument as indicated in
Paragraph 2 of this Operation Section and afterwards
carry out a measurement of a known laser source. Shift the switch on + or – if
you need to increase or decrease the reading displayed on the LCD to match your
reference. This is done by limited pushes on the
ON/Zero
key, until the desired
power value is achieved on the display. Reset the switch to its middle position
CAL
and re-cap. The instrument is ready for new measurements.
6
Calibration
Model: Fit ______ Serial number: ______________
Date of Calibration: _______________
Wavelength 1: _______________nm
Reference Power :________________W Measured Power: _________________W
Wavelength 2: _______________nm
Reference Power :________________W Measured Power: _________________W
LaserPoint certifies that, at the time of calibration, the above listed instrument meets or
exceeds all technical specifications defined at page 8 of this Instruction Manual.
The above listed instrument has been calibrated using standard references traceable to
the National Institute of Standards and Technology (NIST) or Physikalisch Technische
Bundesanstalt (PTB).
This is not a Calibration Certificate for traceability purpose; such a Calibration
Certificate can be quoted on demand.
Important - For Customer Use Only
The reference setting of this instrument can be matched by user to a custom own
reference by means of the instructions reported at Paragraph n.11 of the Operation
Section of this Manual.
3