LaserPoint Fit Скачать руководство пользователя страница 5

  

Operation Section 

 

1-

Safety

 

Take all the required safety procedures needed when 
working with laser beams and wear protection glasses all 
the time! The laser beam must hit ONLY the sensor head 
front side. Avoid to incidentally expose the protected 
stainless steel stem to the laser beam as dangerous back-

reflections can be produced! Never touch the sensor head as this part can reach 
70°C of temperature! 
In the case the laser power density is close to the damage threshold of the head 
absorber coating, damages like coating discoloration or underlying metal melting 
could occur. In both cases consult LaserPoint. 

 

2-Starting the Instrument 

To switch on the instrument, keep the 

ON/Zero

 button 

pushed until the LCD shows the Fit Model (Fit 50, Fit 200, 
Fit 500 shown in 2 steps), then release it (this step takes 

approximately 2 seconds). The instrument will also display one of the 2 available 
laser wavelengths: e.g. CO2. The wavelength displayed the first time is random; 
to change it refer to Paragraph 10 of this section. 

 

 

 

   

 

 

3-Auto-check Cycle 

When the 

ON/Zero

 button is released, the instrument starts 

an auto-check cycle during which both circuit and display 
are controlled and reset. During this cycle, the LCD shows 
all available segments of alphanumeric characters and the 

bar graph at full scale. Upon completion of this cycle, the LCD displays the last 
recorded value of measured power. 

          

 

 

 

 

4-Zeroing/Resetting 

Before starting the first measurement or a new one, the 
instrument needs to be zeroed; to do so, just push and soon 
release the 

ON/Zero

 button. After the LCD has displayed 

CAL

”, the lateral LED turns steady green and 

“FIT”

 

blinks. Any previously recorded value is zeroed and the instrument is ready to 
measure. 
An automatic Zero is performed by Fit after every measurement.

 

 

 

5-Measurement Cycle 

Check the laser shutter is closed; align the probe to the 
laser beam path and start the measurement cycle by opening 
the laser shutter. Avoid taking measurements during laser 
warm up or beam instabilities. 

The measurement starts automatically. During the acquisition time, 

“RUN” 

and the 

LED both blink for approximately 5 seconds. The bar graph indicates the head 
temperature on an arbitrary scale. 
 

6-Measurement Display 

As soon as the measurement is over the LED turns into 
steady red. Remove the probe from the laser beam: the 
display shows the actual power, and the remaining thermal 
capacity. If the bar graph is below full scale, one or more 

measurements can be carried out before the sensor head needs to be cooled. 
Approximately after 10 seconds the LED automatic shuts off. The FIT performs an 
automatic Zero and the LED turns steady green. The LCD shows the last measured 
power value . Return to Par.5 for a new measurement  
                    

7-Cooling the Head

 

Should the probe reach its limit temperature, the LCD  
shows “ 

COOL

” and the LED blinks red. Use 

spontaneous or air forced cooling. 

Never use liquids!!

  

As soon as the temperature returns below its maximum 

allowable limit, the LCD displays the last measured power value. 

Warning

: sudden cool downs and  subsequent strong temperature variations on 

the head may cause a re-start of instrument and spurious measurements. 
 

8-Auto Off /Battery Warning 

When the battery pack is  getting exhausted, its symbol 
pops up on the LCD to  warns that approx. 10 hours of 
residual power remain. 

Note

: to increase the battery life, 

the instrument automatically switches off after 5 minutes 

of stand by operation.  
To retrieve the last recorded value, return to Paragraph 2. After the internal auto 
check cycle, the display will show the last measured value. 
To intentionally shut the instrument off, keep the 

ON/Zero

 button pushed until the 

LCD shuts off (this takes approximately 4 seconds). 
 

                                                                                                                                           5 

 

 

Содержание Fit

Страница 1: ...Nov 2008 Firm vers 11 1 WARNING The user of this instrument must be trained to the use of power lasers and associated risks LaserPoint is in no way liable for any damage resulting from improper careless or above rated limits use of this instrument ...

Страница 2: ...dment 2 1997 Limits and methods of measurement of radio interference characteristics of information technology equipment Class B EN 61000 4 2 1995 Electromagnetic compatibility EMC Part 4 Testing and measurements techniques Section 4 2 Electrostatic discharge Class B EN 61000 4 4 1995 Electromagnetic compatibility EMC Part 4 Testing and measurements techniques Section 4 Electrical fast transient b...

Страница 3: ...technical specifications defined at page 8 of this Instruction Manual The above listed instrument has been calibrated using standard references traceable to the National Institute of Standards and Technology NIST or Physikalisch Technische Bundesanstalt PTB This is not a Calibration Certificate for traceability purpose such a Calibration Certificate can be quoted on demand Important For Customer U...

Страница 4: ...ely 2 seconds The instrument will also display one of the 2 available laser wavelengths e g CO2 The wavelength displayed the first time is random to change it refer to Paragraph 10 of this section 3 Auto check Cycle When the ON Zero button is released the instrument starts an auto check cycle during which both circuit and display are controlled and reset During this cycle the LCD shows all availab...

Страница 5: ... turns steady green The LCD shows the last measured power value Return to Par 5 for a new measurement 7 Cooling the Head Should the probe reach its limit temperature the LCD shows COOL and the LED blinks red Use spontaneous or air forced cooling Never use liquids As soon as the temperature returns below its maximum allowable limit the LCD displays the last measured power value Warning sudden cool ...

Страница 6: ... power measurement carried out at this wavelength setting 11 Customer s calibration This instrument can be re calibrated by the user to match a custom reference Remove the plastic cover of the adjustment switch located at the right side of the instrument body Zero the instrument as indicated in Paragraph 2 of this Operation Section and afterwards carry out a measurement of a known laser source Shi...

Страница 7: ...lf It is emphasised that this no cost return option is only available when you purchase a new product of equivalent type that fulfils the same function Please inform your supplier of your wish to have the old instrument collected when ordering the replacement If you wish to dispose of the instrument without replacing it then the appliance must not be mixed with unsorted municipal waste The crossed...

Страница 8: ...ack 8 Parameter Fit 50 Fit 200 Fit 500 Maximum Measurable Power W 50 200 500 Minimum Measurable Power 3 Accuracy W 1 5 1 5 1 15 1 Absolute Minimum Measurable Power W 1 25 3 7 5 Maximum Laser Beam Diameter mm 20 20 25 Power density damage threshold full scale at YAG wavelength W cm2 10000 Power density damage threshold full scale at CO2 wavelength W cm2 2500 Repeatability 1 ADC Resolution W 0 06 0 ...

Отзывы: