LaserLiner DigiLevel Pro 120 Скачать руководство пользователя страница 4

DigiLevel Pro 30/60/120

Memory functions:

•  With the memory function (6), the current measurement result can be 

saved in the internal memory, whereby the current memory position 
(M1 - M9) is displayed during the saving process (9).

• If more than 9 results are saved, the old ones are overwritten.

•  The memory retrieval button (4) can be pressed to show the most 

recently saved result in the display (8). By repeatingly pressing the 
memory retrieval button (4), all the stored results (9) can be retrieved.

• Return to normal (6)

GB

10

GB

11

Laser radiation!

Do not stare into the beam!

Class 2 laser

< 1 mW · 650 nm

EN 60825-1:2007-10

Warranty:

The warranty is valid for 2 years from the date of purchase. The warranty does 
not cover damage caused by improper use or storage, normal wear and tear, or 
defects which only insigni

fi

 cantly impair the value of the product or its 

functioning. Any tampering by unauthorised persons will render this warranty 
void. In the event that you need to claim warranty, please take the complete 
device together with all information and the invoice to one of our dealers or send 
it in to UMAREX-Laserliner.

 

Electronic measuring
precision: 

+/- 0,1° at 0° and 90°,
+/- 0,2° at 2° – 89°

Display accuracy 0°- 1°

2 decimal places

Display accuracy 1°-90°

1 decimal place

Vial accuracy

+ 0,5 mm/m

Laser class

2 < 1 mW (EN 60825-1:2007-10)

Laser wavelength

650 nm

Working temperature

0°C...40°C (32°F...104°F)

Storage temperature

-20°C ... 65°C

Power supply

2 x 1.5V (Typ AAA/LR03)

Dimensions   

30

60 x 33 x 300 mm

Dimensions   

60

 

60 x 33 x 610 mm

Dimensions 

120

60 x 33 x 1220 mm

Weight 

30/ 60/ 120

0,4 / 0,8   / 1,2 kg

Article-No

. 30/ 60/ 120

081.212A/ 081.210A/ 081.216A 

Subject to technical change without notice 10.2008

Technical data: 

Spirit level function:

•  The alignment to the measuring surface (14) is shown by the vertical and 

horizontal vials (11, 12).

•  Use the vials for optimum vertical and horizontal alignment, and the 

electronic slope angle display for any desired angle of slope.

•  With the aid of the magnets (13), the DigiLevel Pro can be af

fi

 xed to any 

magnetic surface.

Calibration:

• 

(A)

 Place the unit with its measuring edge (14) down on a straight sur-

face and mark the positions of the unit‘s ends on the underlying surface 
(see Fig. below). Switch the unit on (2) and press the 

REF

 button (1) until 

CAL 1

 starts blinking. Shortly thereafter an acoustic signal will sound and 

CAL 2

 will appear in the display. Now turn the unit around horizontally 

(end-for-end) by 180° 

(B)

 such that its ends are positioned opposite where 

they previously were but again exactly at the underlying surface marks 
(reverse measurement). Again press the 

REF

 button (1) until 

CAL 2

 blinks. 

The subsequent acoustic signal concludes the process. 

•  Follow-up test: The unit is properly calibrated when it displays the same 

measurement value, i.e. the underlying surface‘s deviation from absolute 
level, in both positions (0° and 180°).

Switching off:

•  To switch off the DigiLevel, press the on/off switch (2) for at least

3 secs. When not in use, the DigiLevel switches off automatically.

(A)

(B)

    0°

180°

General safety instructions
Attention:

 Do not look directly into the laser beam! The laser may not 

get into the hands of children! Do not point this device unnecessarily 
toward persons.

Note:

 Regularly check the unit‘s calibration before use, after transporting 

and extended periods of storage.