![Larson Davis SoundTrack LxT Скачать руководство пользователя страница 74](http://html1.mh-extra.com/html/larson-davis/soundtrack-lxt/soundtrack-lxt_user-manual_3712247074.webp)
LxT Manual
Overall Tab
5-19
C minus A and Impulsivity
Figure 5-22 shows the
C minus A
level and
Impulsivity
pages.
FIGURE 5-22 C-A Level and Impulsivity Page
The integrated levels for L
Aeq
are always calculated using
the linear detector, regardless of the value selected in the
SLM Setup. The LAIeq value is from the impulse detector.
L
CSeq
and L
ASeq
are equivalent levels measured using Slow
RMS averaging and using frequency weightings of C and A,
respectively. The difference between them, L
CSeq -
L
ASeq,
is
often used as an indicator of the amount of low frequency
content in sounds.
It is also used as a parameter for the selection of hearing
protectors, since noise fields having large amounts of low
frequency sound can require more effective hearing
protectors than would otherwise be indicated by the
measured L
ASeq
level alone.
The parameter LAFTMS only appears when
Takt Maximal
Data has been selected on the
Preferences
tab.
Содержание SoundTrack LxT
Страница 1: ...LxT Manual for SoundTrack LxT SoundExpert LxT TM...
Страница 55: ...4 18 Measurement Settings Tabs LxT Manual...
Страница 81: ...5 26 View Spectrum Normalized Optional LxT Manual...
Страница 93: ...6 12 Data Storage After Improper Shutdown LxT Manual...
Страница 110: ...LxT Manual Certification 7 17...
Страница 143: ...11 6 Control Panel Data Explorer LxT Manual...
Страница 167: ...12 24 Options LxT Manual...
Страница 179: ...14 4 User LxT Manual FIGURE 14 5 shows the User tab FIGURE 14 5 User Tab...
Страница 185: ...15 6 File System LxT Manual...
Страница 191: ...16 6 Miscellaneous Parameters LxT Manual...
Страница 206: ...LxT Manual Technical Specifications A 9 Typical Z Weight Frequency Response...
Страница 259: ...A 62 Technical Specifications LxT Manual...